Why the study?
What are the mortality and complication rates associated with permanent endocardial pacing?
What are the mortality and complication rates associated with permanent endocardial pacing?
In an early cohort from 1970-1974, permanent endocardial pacing was associated with a 20% rate of electrode complications and a mean battery life of 26 months, highlighting significant early technical limitations.
Early pacing showed high electrode complications; leaves open assessment of modern device durability.
Primary results and follow-up observations in 90 patients with permanent endocardial pacing--covering the period from 1970 to June 1974--are reported. The primary mortality was 3% (3/90), and mortality 4% (4/90). Electrode complications were the most common problem and occurred in 20% implantations. Early or late dislocation of the electrode tip was seen in 12%. With the improved technique is was possible to decrease it significantly through the period. The mean battery life time of Siemens-Elema pacemakers was 26 months. 10% of the failed batteries were replaced urgently. So far the rather high frequency of electrode complications diminishes the advantages of the pacemaker treatment.