Key result
Endocardial pacemaker electrodes allowed for long-term complication-free pacing, with 40% of patients surviving 10 or more years and 62% of these survivors experiencing no electrode problems.
Why the study?
Do endocardial pacemaker electrodes provide long-term complication-free pacing in patients requiring pacemakers?
Cohort (n=328)
Do endocardial pacemaker electrodes provide long-term complication-free pacing in patients requiring pacemakers?
Endocardial pacemaker electrodes can provide stable, complication-free pacing over a 10-year period with consistent stimulation thresholds.
Answers are grounded in this paper and cite their sources.
Supports durability in long-term survivors; leaves open generalizability and need for randomized comparisons.
Epicardial electrodes implanted in 20 patients between 1959 and 1962 had a short-term effectiveness. A change to endocardial stimulation was later necessary in 15 of these patients. Endocardial electrodes inserted between 1962 and 1967 in 306 patients were included in this study; 121 (40%) of these patients survived for 10 years or more. Seventy-five (62%) of the long-term survivors had no problems with their electrodes. In 39 of these patients the stimulation thresholds were measured after both 5 and 10 years and found to be 2.3 +/- 0.6 volts and 2.2 +/- 0.8 volts, respectively. Hence, long-term complication-free endocardial pacing can be accomplished.
No takes yet. Share an insight, caveat, or question.
Lagergren et al. (1979) conducted a cohort in Pacemaker implantation (n=328). Endocardial pacemaker electrodes vs. Epicardial electrodes was evaluated on 10-year survival. Endocardial pacemaker electrodes allowed for long-term complication-free pacing, with 40% of patients surviving 10 or more years and 62% of these survivors experiencing no electrode problems.
Synapse has enriched 10 related papers. Consider them for comparative context: