This paper explores Rayleigh scattering thermometry via a wavelength of 355 nm through a unique measurement scheme. In this context, the p-polarization and s-polarization Rayleigh scattering of flame and air (as the temperature calibration reference) are measured. Subtraction of p-polarization Rayleigh scattering intensity from that of s-polarization is proposed to eliminate the background noise and fluorescence interference influence to reduce the temperature measurement uncertainties. To validate this method, the temperature field of CH4/N2/O2 premixed flame at ϕ=0.78 on a McKenna burner is detected by this Rayleigh scattering thermometry, and the axial temperature profile is validated with the literature data. Within the region of interest domain (−5mm to 5 mm in the radial direction), an overall temperature measurement system precision of ±46.5K is reported. The influence of both p-polarization Rayleigh scattering and laser sheet inhomogeneity on the temperature measurement is further quantitatively studied. The measurement uncertainties relevant to laser energy variation and flame Rayleigh scattering cross-section variation due to temperature increase are specified as 1.4% and 2%–8%, respectively. Eventually, temperature measurements of single-shot images are attempted, and the large signal dynamic range (100–1000 [a.u.]) indicates a promising potential for temperature field interpretation of turbulence combustion.
No takes yet. Share an insight, caveat, or question.
Wang et al. (2019) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: