Why the study?
Does fully automated preprocedural MSCT segmentation reduce contrast medium and radiation exposure during TAVR compared to standard intraprocedural angiographic positioning?
Does fully automated preprocedural MSCT segmentation reduce contrast medium and radiation exposure during TAVR compared to standard intraprocedural angiographic positioning?
Fully automated preprocedural MSCT segmentation provides accurate C-arm angulation prediction for TAVR, significantly reducing intraprocedural contrast and radiation exposure.
Supports automated MSCT for TAVR angulation prediction; hypothesis-generating and requires randomized confirmation before practice change.
BACKGROUND: Preprocedural manual multi-slice-CT-segmentation tools (MSCT-ST) define the gold standard for planning transcatheter aortic valve replacement (TAVR). They are able to predict the perpendicular line of the aortic annulus (PPL) and to indicate the corresponding C-arm angulation (CAA). Fully automated planning-tools and their clinical relevance have not been systematically evaluated in a real world setting so far. METHODS AND RESULTS: The study population consists of an all-comers cohort of 160 consecutive TAVR patients with a drop out of 35 patients for technical and anatomical reasons. 125 TAVR patients underwent preprocedural analysis by manual (M-MSCT) and fully automated MSCT-ST (A-MSCT). Method-comparison was performed for 105 patients (Cohort A). In Cohort A, CAA was defined for each patient, and accordance within 10° between M-MSCT and A-MSCT was considered adequate for concept-proof (95% in LAO/RAO; 94% in CRAN/CAUD). Intraprocedural CAA was defined by repetitive angiograms without utilizing the preprocedural measurements. In Cohort B, intraprocedural CAA was established with the use of A-MSCT (20 patients). Using preprocedural A-MSCT to indicate the corresponding CAA, the levels of contrast medium (ml) and radiation exposure (cine runs) were reduced in Cohort B compared to Cohort A significantly (23.3±10.3 vs. 35.3 ±21.1 ml, p = 0.02; 1.6±0.7 vs. 2.4±1.4 cine runs; p = 0.02) and trends towards more safety in valve-positioning could be demonstrated. CONCLUSIONS: A-MSCT-analysis provides precise preprocedural information on CAA for optimal visualization of the aortic annulus compared to the M-MSCT gold standard. Intraprocedural application of this information during TAVR significantly reduces the levels of contrast and radiation exposure. TRIAL REGISTRATION: ClinicalTrials.gov NCT01805739.
No takes yet. Share an insight, caveat, or question.
Veulemans et al. (2016) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: