Answers are grounded in this paper and cite their sources.
This paper describes an automated method for counting and sizing fish eggs in the diameter range 200–2500 μm. Eggs, suspended in water, are pumped at a controlled rate through an electronic sensor which produces a voltage pulse which is proportional in amplitude to particle size. The sensor is connected to a HIAC Criterion PC‐320 particle analyser which is in turn connected to a Tracor Northern pulse height analyser. This equipment separates the voltage pulses into 398 recording channels. An ACT 1 Sirius microcomputer is used to store the data, transform it to egg size and then produce a histogram of egg size distribution in 20‐μm class intervals. The method has a counting accuracy of 99.5 ± 0.5%. It has been compared with two manual methods of estimating egg numbers.
No takes yet. Share an insight, caveat, or question.
Witthames et al. (1987) studied this question.