Why the study?
Guidelines recommend blood pressure measurement on a bare arm, which is not always easy, creating a need to synthesize evidence on the effect of a sleeve.
Does measuring blood pressure over a sleeve or below a rolled-up sleeve affect the recorded values compared to measuring on a bare arm?
Does measuring blood pressure over a sleeve or below a rolled-up sleeve affect the recorded values compared to measuring on a bare arm?
Measuring blood pressure over a thin sleeve does not significantly impact recorded values and is preferable to rolling up the sleeve, which may cause a larger overestimation.
Answers are grounded in this paper and cite their sources.
Office BP over thin sleeves shows negligible bias; leaves open whether guidelines should relax bare-arm requirements pending higher-quality data.
OBJECTIVE: Several guidelines call for blood pressure (BP) measurement on a bare arm, which is not always easy. This systematic review aims to synthesize existing evidence concerning the effect of a sleeve on BP measurement. METHODS: Pubmed and Embase were searched for cross-sectional studies comparing BP values measured on a bare arm, over a sleeve or below a rolled-up sleeve. A meta-analysis was conducted on available data. RESULTS: Thirteen articles were selected from 720 references. All studies reported office BP values, 12 compared measurements on a bare arm and on a sleeve, and four also performed measurements below a rolled-up sleeve, with heterogeneous sleeve types and thicknesses. Most studies had a high risk of bias. Three studies showed a small overestimation of BP measured over a sleeve, but the remaining 10 studies did not find statistically significant differences between measurements. Meta-analysis showed a nonsignificant 0.59 mmHg [95% confidence interval (CI) -0.11 to +1.30; P = 0.10] overestimation of SBP measured over a sleeve when the thinnest sleeve was considered for studies that investigated various thicknesses, a nonsignificant 1.10 mmHg (95% CI -0.21 to +2.40; P = 0.10) overestimation of SBP when the thickest sleeve was considered, and a nonsignificant 2.76 mmHg (95% CI -0.96 to +6.47; P = 0.15) overestimation of SBP measured below a rolled-up sleeve. CONCLUSION: Measuring BP over a thick sleeve in the office may result in a small overestimation of recorded values but measuring over a thin sleeve does not appear to have a significant impact and, in any case, should be preferred to rolling it up.
No takes yet. Share an insight, caveat, or question.
Seguret et al. (2020) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: