Answers are grounded in this paper and cite their sources.
Detection of the t-point of the ICA curve may allow for sensorless assessment of cardiac output in LVAD-supported conditions.
It is the goal of this section to publish material that provides information regarding specific issues, aspects of artificial organ application, approach, philosophy, suggestions, and/or thoughts for the future Abstract: In this series, we investigated the meaning of the t‐point of index of motor current amplitude (ICA) curve from a point of view of flow rate on in vitro and in vivo studies. On mock circulation loop and left ventricular assist device (LVAD)–equipped pigs, we detected the t‐point and compared the pump flow at the t‐point with the simultaneous cardiac output. The pump flow at the t‐point showed high correlation against the simultaneous cardiac output for in vitro or in vivo study. By detection of the t‐point of the ICA curve and measuring or estimating the pump flow at t‐point, the cardiac output may be assessed without any sensor in various cardiac conditions.
No takes yet. Share an insight, caveat, or question.
Endo et al. (2003) studied this question.