Key result
Motor current amplitude of a centrifugal blood pump showed an excellent correlation with left ventricular pressure in both in vitro (r=0.77-0.99) and in vivo sheep models (r=0.70-0.94).
Why the study?
Does motor current amplitude monitoring correlate with left ventricular pressure in a centrifugal blood pump model?
Does motor current amplitude monitoring correlate with left ventricular pressure in a centrifugal blood pump model?
Motor current amplitude from a centrifugal blood pump correlates well with left ventricular pressure, suggesting it could be a useful index for monitoring native cardiac function during VAD support.
No takes yet. Share an insight, caveat, or question.
May enable noninvasive LV pressure monitoring during centrifugal VAD support; leaves open human validation.
Daiki Kikugawa (2001) studied Ventricular assist device monitoring (n=5). Motor current amplitude monitoring of a centrifugal blood pump was evaluated on Correlation between motor current and left ventricular pressure. Motor current amplitude of a centrifugal blood pump showed an excellent correlation with left ventricular pressure in both in vitro (r=0.77-0.99) and in vivo sheep models (r=0.70-0.94).