The properties of three Ge(Li) detectors as γ-ray spectrometers were measured over the temperature range 5.0 to 170°K. These detectors, selected to show a range of charge-trapping effects, all exhibited predominantly electron trapping, and optimum resolution occurred between 20 and 30°K where electron-trapping effects were a minimum. As the temperature was reduced it was possible to increase the operating voltage, giving some improvement in resolution. Below 20°K the γ-ray resolution rapidly deteriorated due to the appearance of a slow-rising component (up to 20 μsec) in the preamplifier pulses.