Key result
The topographic cellular contour extraction algorithm for 3-D right atrial reconstruction yielded an average volume quantification error of 8% ± 5% compared to manual delineation.
Why the study?
Does the TCCE algorithm accurately quantify right atrial volume from transesophageal echocardiography compared to manual delineation?
Does the TCCE algorithm accurately quantify right atrial volume from transesophageal echocardiography compared to manual delineation?
The TCCE algorithm provides automated, online 3-D reconstruction of the right atrium from echocardiography with acceptable accuracy compared to manual delineation, potentially aiding in procedures like ASD closure.
May support automated right atrial volumetry from echocardiography; leaves open prospective clinical validation before adoption.
A computational method providing online, automated 3-D reconstruction of the right atrium of the human heart is presented in this paper. Endocardial boundaries were extracted from transesophageal ultrasound data by a topographic cellular contour extraction (TCCE) algorithm. The TCCE method was implemented on a continuous-time, analog, massively parallel processor, and on a digital serial processor. Processing speeds were 500 or 40 frames per second, depending on the hardware used. Extracted boundary point sets were rendered into a 3-D mesh and the volume of the cavity was quantified. Accuracy of volume quantification was validated on 60 in vitro static phantoms and 12 clinical subjects. For the clinical recordings, reference volumes were estimated from manually delineated endocardial boundaries. The average error of volume quantification by the TCCE method was 8% +/-5% ( n = 12), the average of the interobserver variability between two independent human experts was 5% +/-2% ( n = 6). Interactive planning of atrial septal defect closure in pediatric cardiology is presented as an example that demonstrates a potential clinical application of the method.
No takes yet. Share an insight, caveat, or question.
Hillier et al. (2009) studied Right atrium imaging (n=12). Topographic cellular contour extraction (TCCE) algorithm vs. Manual delineation was evaluated on Average error of volume quantification. The topographic cellular contour extraction algorithm for 3-D right atrial reconstruction yielded an average volume quantification error of 8% ± 5% compared to manual delineation.