Key result
The compressed sensing method with a compression ratio of 8 reduced energy consumption to 2.32 J compared to 2.65 J without compression, while maintaining an acceptable signal reconstruction quality.
Why the study?
To develop an efficient method to reduce transmitted data and increase battery life in Wi-Fi IoT devices for continuous ECG monitoring.
Answers are grounded in this paper and cite their sources.
A novel compressed sensing method for Wi-Fi IoT ECG monitoring reduces energy consumption while maintaining acceptable signal reconstruction quality.
The paper presents an Internet of Things (IoT) prototype which consists of a data acquisition device wirelessly connected to Internet via Wi-Fi, for continuous electrocardiogram (ECG) monitoring. The proposed system performs a novel Compressed Sensing (CS) based method on ECG signal with the aim of reducing the amount of transmitted data, thus realizing an efficient way to increase the battery life of such devices. For the assessment of the energy consumption of the device, an experimental setup was arranged and its description is presented. The evaluation of the reconstruction quality of the ECG signal in terms of Percentage of Root-mean-squared Difference (PRD) is reported for several Compression Ratios (CRs). The obtained experimental results clearly demonstrate the robustness and usefulness of the Wi-Fi based IoT devices adopting the considered CS-method for data compression of ECG signals. Furthermore, it allows reducing the energy consumption of the IoT device, by increasing the CR, without significantly degrading the quality of the reconstructed ECG signal.
No takes yet. Share an insight, caveat, or question.
Balestrieri et al. (2020) studied ECG monitoring. Compressed sensing method for ECG data transmission vs. No compression was evaluated on Energy consumption (Joules) for 40 seconds at Compression Ratio (CR) = 8. The compressed sensing method with a compression ratio of 8 reduced energy consumption to 2.32 J compared to 2.65 J without compression, while maintaining an acceptable signal reconstruction quality.
Synapse has enriched 10 related papers. Consider them for comparative context: