Answers are grounded in this paper and cite their sources.
A multiplexed diode-laser sensor system comprising two diode lasers and fiber-optic components has been developed to nonintrusively monitor temperature and species mole fraction over a single path using both scanned-and fixed-wavelength laser absorption spectroscopy techniques. In the scanned-wavelength method, two InGaAsP lasers were current tuned at a 2-kHz rate across H 2 O transitions near 1343 nm and 1392 nm in the 2v 1 and v 1 + v 3 bands. Gas temperature was determined from the ratio of single-sweep integrated line intensities. Species mole fraction was determined from the measured line intensity and the calculated line strength at the measured temperature. In the fixed-wavelength method, the wavelength of each laser was fixed near the peak of each absorption feature using a computer-controlled laser line-locking scheme. Rapid measurements of gas temperature were obtained from the determination of peak line-intensity ratios. The system was applied to measure temperature and species concentration in the postflame gases of an H 2 -O 2 flame. The good agreement between the laser-based measurements obtained using scanned- and fixed-wavelength methods with those recorded with thermocouples demonstrates the flexibility and utility of the multiplexed diode-laser sensor system and the potential for rapid, continuous measurements of gasdynamic parameters in high-speed or transient flows with difficult optical access.
No takes yet. Share an insight, caveat, or question.
Baer et al. (1996) studied this question.