Near infrared region (NIR) spectrophotometry was used to determine concentrations of glucose, fructose and sucrose in aqueous mixtures at 10, 25, and 40% total sugars. Calibration equations and predicted individual concentrations in each total sugar level are reported. The standard errors of prediction varied between 0.35 and 0.69 and confidence limits ranged from ± 0.7 to ± 1.1%. In developing a single equation for predicting individual sugars in solutions, suitable for use over a range of total sugar concentrations, spectral data needed correction for sample density differences. However, at total sugar concentrations above 25%, accompanying wavelength shifts were of sufficient magnitude that a satisfactory single equation could not be developed. Equations for predicting individual sugars in solutions ranging from 10–25% total sugar concentrations were developed. For higher values of total sugars, appropriate calibrations were necessary.