Key result
Continuous 12-lead Holter recordings provided a more accurate reflection of 24-hour cardiac electrical activity than resting ECGs, which revealed sex differences (e.g., 20 ms longer QTcF in women).
Why the study?
Resting 12-lead ECGs in clinical trials are limited to discrete timepoints and may miss valuable electrical activity, leading to increased use of continuous Holter recorders, but comparative data between the two approaches are lacking.
Does continuous 12-lead Holter recording provide consistent electrocardiogram interval measurements compared to standard 12-lead resting ECGs in healthy subjects?
RCT (n=40)
double-blind
randomized crossover
Does continuous 12-lead Holter recording provide consistent electrocardiogram interval measurements compared to standard 12-lead resting ECGs in healthy subjects?
Continuous 12-lead Holter monitoring provides an accurate reflection of 24-hour cardiac electrical activity comparable to resting ECGs, though manual adjudication remains necessary.
Answers are grounded in this paper and cite their sources.
Continuous Holter better captures 24-hour ECG dynamics in trials than sparse resting ECGs; extends randomized evidence for its routine adoption in cardiac safety assessments.
In clinical trials, traditionally only a limited number of 12-lead resting electrocardiograms (ECGs) can be recorded and, thus, long intervals may elapse between assessment timepoints and valuable information may be missed during times when patients' cardiac electrical activity is not being monitored. These limitations have led to the increasing use of Holter recorders which provide continuous data registrations while reducing the burden on patients and freeing up time for clinical trial staff to perform other tasks. However, there is a shortage of data comparing the two approaches. In this study, data from a randomized, double-blind, four-period, crossover thorough QT study in 40 healthy subjects were used to compare continuous 12-lead Holter recordings to standard 12-lead resting ECGs which were recorded in parallel. Heart rate and QT interval data were estimated by averaging three consecutive heartbeats. Values exceeding the sample average by more than 5% were tagged as outliers and excluded from the analysis. Visual comparisons of the ECG waveforms of the Holter signal showed a good correlation with resting ECGs at matching timepoints. Resting ECG data revealed sex differences that Holter data did not show. Specifically, women were found to have a longer QTcF of 20 ms, while men had a lower heart rate. We found that continuous recordings provided a more accurate reflection of changes in cardiac electrical activity over 24 hr. However, manual adjudication is still required to ensure the quality and accuracy of ECG data, and that only artifacts are removed thereby avoiding loss of true signals.
No takes yet. Share an insight, caveat, or question.
Mendzelevski et al. (2021) conducted an RCT in healthy subjects (n=40). Continuous 12-lead Holter recordings vs. Standard 12-lead resting ECGs was evaluated on Heart rate and QT interval data. Continuous 12-lead Holter recordings provided a more accurate reflection of 24-hour cardiac electrical activity than resting ECGs, which revealed sex differences (e.g., 20 ms longer QTcF in women).