Key result
Inappropriate ICSD therapy occurred in 7.6% of patients, with 76% of recurrences happening within 180 days, suggesting that the interval for driving restrictions can be safely reduced.
Observational (n=1,570)
Yes
The majority of recurrent inappropriate ICSD therapies occur within 180 days, suggesting that the duration of driving restrictions after an inappropriate shock could potentially be reduced.
Answers are grounded in this paper and cite their sources.
May support shorter driving restrictions after inappropriate ICSD therapy; hypothesis-generating and requires prospective validation before practice change.
BACKGROUND: Little is known regarding the appropriate duration for driving restrictions after inappropriate implantable cardiac shock device (ICSD) therapy. METHODS AND RESULTS: We evaluated the Nippon Storm Study data, and found that inappropriate ICSD therapy occurred in 114 (7.6%) patients during a median follow-up of 464 days. Among those patients, 25 experienced further inappropriate ICSD therapy during a subsequent median follow-up of 380 days. Time-dependent recurrence of inappropriate ICSD therapy occurred in 19 (76%) patients within 180 days. CONCLUSIONS: The interval for driving restrictions after inappropriate ICSD therapy can be reduced.
No takes yet. Share an insight, caveat, or question.
Noda et al. (2014) conducted an observational in Implantable cardiac shock device (ICSD) recipients (n=1,570). Inappropriate ICSD therapy was evaluated on Incidence of first inappropriate ICSD therapy. Inappropriate ICSD therapy occurred in 7.6% of patients, with 76% of recurrences happening within 180 days, suggesting that the interval for driving restrictions can be safely reduced.
Synapse has enriched 10 related papers. Consider them for comparative context: