Key result
The proposed artifact removal method eliminated 96% to 100% of coarse artifacts and completely reduced slow-changing artifacts for capacitive ECG signals larger than 0.3 V.
Why the study?
Capacitive electrodes preserve driver comfort during ECG monitoring, but the recorded signals have large artifacts whose spectral bands overlap with the ECG, limiting conventional filtering.
A novel artifact removal method based on linear trend fluctuation estimation effectively reduces coarse and slow-changing artifacts in capacitive ECG signals during movement.
Answers are grounded in this paper and cite their sources.
Proposed cECG artifact reduction may aid driver monitoring; leaves open real-world validation before clinical use.
The development of smart cars with e-health services allows monitoring of the health condition of the driver. Driver comfort is preserved by the use of capacitive electrodes, but the recorded signal is characterized by large artifacts. This paper proposes a method for reducing artifacts from the ECG signal recorded by capacitive electrodes (cECG) in moving subjects. Two dominant artifact types are coarse and slow-changing artifacts. Slow-changing artifacts removal by classical filtering is not feasible as the spectral bands of artifacts and cECG overlap, mostly in the band from 0.5 to 15 Hz. We developed a method for artifact removal, based on estimating the fluctuation around linear trend, for both artifact types, including a condition for determining the presence of coarse artifacts. The method was validated on cECG recorded while driving, with the artifacts predominantly due to the movements, as well as on cECG recorded while lying, where the movements were performed according to a predefined protocol. The proposed method eliminates 96% to 100% of the coarse artifacts, while the slow-changing artifacts are completely reduced for the recorded cECG signals larger than 0.3 V. The obtained results are in accordance with the opinion of medical experts. The method is intended for reliable extraction of cardiovascular parameters to monitor driver fatigue status.
No takes yet. Share an insight, caveat, or question.
Tamara Škorić (2021) studied Moving and driving subjects. Artifact removal method based on estimating fluctuation around linear trend was evaluated on Elimination of coarse and slow-changing artifacts. The proposed artifact removal method eliminated 96% to 100% of coarse artifacts and completely reduced slow-changing artifacts for capacitive ECG signals larger than 0.3 V.