Key result
Apple Watch ECGs demonstrated moderate to strong agreement with lead 1 of a 12-lead ECG for basic intervals in healthy adults, including 100% concordance for sinus rhythm interpretation.
Why the study?
The Apple Watch Series 4 can detect atrial fibrillation and record a single-lead ECG, but the clinical accuracy of its waveforms compared to lead 1 of a 12-lead ECG was unclear.
Does the Apple Watch Series 4 single-lead ECG accurately measure basic ECG intervals compared to lead 1 of a 12-lead ECG in healthy adults?
Observational (n=43)
No
Does the Apple Watch Series 4 single-lead ECG accurately measure basic ECG intervals compared to lead 1 of a 12-lead ECG in healthy adults?
The Apple Watch Series 4 produces accurate single-lead ECGs with moderate to strong agreement in basic ECG intervals compared to lead 1 of a standard 12-lead ECG in healthy adults.
Apple Watch ECG shows moderate-to-strong interval agreement in healthy adults; leaves open validation in patients with arrhythmias before clinical adoption.
Background: The Apple Watch Series 4 (AW) can detect atrial fibrillation and perform a single-lead electrocardiogram (ECG), but the clinical accuracy of AW ECG waveforms compared to lead 1 of a 12-lead ECG is unclear. Objective: The purpose of this study was to assess the accuracy of interval measurements on AW ECG tracings in comparison to lead 1 on a 12-lead ECG. Methods: We obtained ECGs at a university hospital of healthy volunteers age >18 years. ECG waveforms were measured with calipers to the nearest 0.25 mm. When possible, 3 consecutive waveforms in lead 1 were measured. Waveform properties, including intervals, were recorded. Concordance correlation coefficients and Bland-Altman plots were used to assess level of agreement between devices. Results: Twelve-lead (n = 113) and AW (n = 129) ECG waveforms from 43 volunteers (mean age 31 years; 65% female) were analyzed. Sinus rhythm interpretation between devices was 100% concordant. No arrhythmias were recorded. Mean difference (d) for heart rate was 1.16 ± 4.33 bpm (r = 0.94); 3.83 ± 113.54 ms for RR interval (r = 0.79); 5.43 ± 17 ms for PR interval (r = 0.83); -6.89 ± 14.81 ms for QRS interval (r = 0.65); -11.27 ± 22.9 ms for QT interval (r = 0.79); and -11.67 ± 27 ms for QTc interval (r = 0.57). There was moderate (d <40 ms) to strong (d <20 ms or < 5 bpm) agreement between devices represented by Bland-Altman plots. Conclusion: The AW produces accurate ECGs in healthy adults with moderate to strong agreement of basic ECG intervals.
No takes yet. Share an insight, caveat, or question.
Saghir et al. (2020) conducted an observational in Healthy adults (n=43). Apple Watch Series 4 ECG vs. Lead 1 of a 12-lead ECG was evaluated on Accuracy of interval measurements (heart rate, RR, PR, QRS, QT, QTc) and sinus rhythm interpretation. Apple Watch ECGs demonstrated moderate to strong agreement with lead 1 of a 12-lead ECG for basic intervals in healthy adults, including 100% concordance for sinus rhythm interpretation.