May 1, 1974CirculationOpen Access

Patterns of Ventricular Activity During Catheter Defibrillation

View Full Paper
Ask AI
Bookmark
Share

Why the study?

Does catheter defibrillation require total depolarization of the entire myocardium in a canine model?

Population

39 dogs undergoing 120 catheter fibrillation-defibrillation episodes

Design

Preclinical

Authors

MMMorton M. MowerMMM. MirowskiJSJoseph F. Spear

Discussion

Loading...

Member takes

Overview

May not require total myocardial depolarization for defibrillation in canines; leaves open translation to human catheter use.

Structured PICO

Does catheter defibrillation require total depolarization of the entire myocardium in a canine model?

P
Population
39 dogs undergoing 120 catheter fibrillation-defibrillation episodes
I
Intervention
Catheter defibrillation
O
Outcome
Electrocardiogram patterns during defibrillationsurrogate

Catheter defibrillation may not require simultaneous depolarization of the entire myocardium, as evidenced by distinct ECG patterns during the procedure.

Cite This Study

Mower et al. (1974) studied this question.

synapsesocial.com/papers/6a925d9869d1d3588867644ahttps://doi.org/10.1161/01.cir.49.5.858
View Full Paper
Ask AI
Bookmark
Share

Also Consider

Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context:

  1. 1Ventricular Defibrillation in the Dog With a Bielectrode Intravascular Catheter1973 · 36 citations
  2. 2Transthoracic Ventricular Defibrillation with Triangular and Trapezoidal Waveforms1966 · 117 citations
  3. 3Feasibility and Effectiveness of Low-Energy Catheter Defibrillation in Man1973 · 151 citations
  4. 4The Development of the Transvenous Automatic Defibrillator1972 · 89 citations
  5. 5CLOSED CHEST DEFIBRILLATION OF THE HEART1958 · 72 citations