This is the authors' abstract. We don't add key points for this paper.
We describe the design and working of a computer-controlled, closed anaesthetic breathing system which rapidly achieves and maintains a prescribed end-tidal concentration of isoflurane in oxygen. The system is simple to set up and not expensive; the only nonstandard component is a modified glass syringe. We have demonstrated that gas analysers may contribute as much as the patient to the accumulation of nitrogen within the breathing system. Details of our clinical experience with the system are presented in an accompanying article.
Answers are grounded in this paper and cite their sources.
No takes yet. Share an insight, caveat, or question.
LOCKWOOD et al. (1993) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: