Key result
A theoretical model of the CW Doppler ultrasonic flowmeter demonstrates that measurement of blood flow velocity corresponds to estimating the average frequency of the Doppler power spectrum.
Provides a detailed theoretical understanding of the CW Doppler ultrasonic flowmeter by modeling back-scattered ultrasound as a Gaussian random process.
Answers are grounded in this paper and cite their sources.
Theoretical model supports frequency estimation for CW Doppler velocity; leaves open in vivo validation before clinical adoption.
The widespread application of ultrasonic techniques for the measurement of pulsatile blood flow has been hampered by the lack of a detailed theoretical understanding of the Doppler ultrasonic flowmeter. A general model for the Doppler flowmeter based upon stochastic considerations of the scattering of ultrasound by blood is introduced in this paper. The model characterizes the back- scattered ultrasound as a Gaussian random process and the expression for the autocovariance function is derived. For the CW Doppler flowmeter, the power spectral density function is computed, and its significance is emphasized: measurement of blood flow velocity corresponds to estimation of the average frequency of the Doppler power spectrum.
No takes yet. Share an insight, caveat, or question.
Brody et al. (1974) studied this question. Theoretical model of CW Doppler Ultrasonic Flowmeter was evaluated on Autocovariance function and power spectral density function. A theoretical model of the CW Doppler ultrasonic flowmeter demonstrates that measurement of blood flow velocity corresponds to estimating the average frequency of the Doppler power spectrum.
Synapse has enriched 10 related papers. Consider them for comparative context: