Answers are grounded in this paper and cite their sources.
We describe a method for the determination of retinol, alpha-tocopherol, and beta-carotene in serum, using a liquid-chromatographic separation with wavelength-programmed ultraviolet/visible absorbance and amperometric electrochemical detection with a glassy carbon electrode. After protein denaturation and addition of an internal standard, tocol, 250-microL samples are twice extracted with hexane. The reversed-phase, gradient-elution chromatographic separation provides baseline resolution of: the all-trans isomer of retinol from the cis isomers, alpha- from gamma-tocopherol, and all-trans-beta-carotene from alpha-carotene and from cis-beta-carotene isomers. The linearity of response and the detection limits for the two detectors for the three analytes are measured. A comparison of the values obtained for serum extracts shows good agreement between the absorbance and electrochemical detectors.
No takes yet. Share an insight, caveat, or question.
MacCrehan et al. (1987) studied this question.