Key result
The start point of the pulse wave measured by the novel piezo film and conductive textile sensor showed a significant correlation with the reference PP-1000 device (r=0.691, P=0.00).
Effect estimate: r=0.691
p-value: p=0.00
A newly developed portable sensor using piezoelectric film and conductive textile showed reliable correlation with a reference device for measuring pulse wave velocity, suggesting potential for convenient arterial stiffness monitoring.
Supports portable PWV sensor development; leaves open clinical utility pending larger validation studies.
Arterial stiffness is causing the serious problems for human who is suffered from hypertension and metabolic syndrome. So it is important that measure the arterial stiffness for early prevention. Many researches point out that pulse wave velocity(PWV) is the reliable and simple method to predict arterial stiffness. In this paper, we developed the sensing parts that detect the pulse wave and ECG by using piezoelectric film and conductive textile with elastic band. Our system could detect 3ch pulse wave and ECG. Simultaneously, our algorithm extracts the features for calculating the delays among pulse waves. The delays are the significant parameter to estimate PWV, thus we design the experiment for evaluating the performance of our sensing parts. The reference is PP-1000(HanByul Meditech, Korea) that is good for performance evaluation. As a result, the start point of the pulse wave was the most reliable feature for comparing with PP-1000(r=0.691, P=0.00). The results between two operators showed that there is only a slight difference in the reproducibility of the devices. In conclusion, we assume that the suggested sensor could be more comfortable and faithful method for arterial stiffness.
No takes yet. Share an insight, caveat, or question.
Kim et al. (2012) studied this question. Piezo film and conductive textile sensor vs. PP-1000 reference device was evaluated on Correlation of pulse wave start point (SP) with reference device (r=0.691, p=0.00). The start point of the pulse wave measured by the novel piezo film and conductive textile sensor showed a significant correlation with the reference PP-1000 device (r=0.691, P=0.00).
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: