Key result
Wearable distributed-deflection sensor successfully records radial and carotid pulse waveforms matching expected physiological variations.
Answers are grounded in this paper and cite their sources.
A novel wearable distributed-deflection sensor can successfully measure arterial pulse waveforms and derive physiologically relevant tonometric parameters like PWV and AI.
This paper presents a performance investigation of a wearable distributed-deflection sensor in arterial pulse waveform measurement. Built on a flexible substrate, the sensor entails a polymer microstructure embedded with an electrolyte-enabled resistive transducer array. By pressing the sensor against an artery with hold-down pressure exerted by two fingers, the pulse signal from the artery deflects the microstructure and registers as a resistance change by the transducer at the artery site. The radial and carotid pulse signals of five subjects are recorded via the sensor. Related signal-processing algorithms are written in Matlab to remove motion artifacts from a recorded pulse signal, extract its key tonometric parameters, and calculate the Pulse Wave Velocity (PWV) and radial and carotid Augmentation Index (AI). Whereas the tonometric parameters of the measured radial and carotid pulse waveforms on each subject are consistent with the related findings in the literature, the difference in tonometric parameters among the subjects also reveals physiological significance. The measured PWV and radial and carotid AI of the subjects show good agreement with how they should vary with age, gender and hypertension. Finally, the effect of the hold-down pressure on a measured pulse signal and the repeatability of the sensor are examined.
No takes yet. Share an insight, caveat, or question.
Wang et al. (2017) studied this question. Wearable distributed-deflection sensor was evaluated on Radial and carotid pulse signals, Pulse Wave Velocity (PWV), and Augmentation Index (AI). A wearable distributed-deflection sensor successfully recorded radial and carotid pulse waveforms in 5 subjects, with derived tonometric parameters showing good agreement with expected physiological variations.
Synapse has enriched 10 related papers. Consider them for comparative context: