Key result
Precordial pacing in normal canine hearts demonstrated a high safety factor, requiring an average of 12.6 times the pacing current to induce ventricular fibrillation.
Precordial pacing demonstrates a wide safety margin in a normal canine model, requiring approximately 12 times the pacing current to induce ventricular fibrillation.
Answers are grounded in this paper and cite their sources.
Precordial pacing shows high safety margin in normal canines; leaves open translation to human clinical practice.
Temporary cardiac pacing by means of rapidly applied, precordial electrodes can be of great value in selected patients suffering cardiac arrest or profound bradycardia, post-defibrillation atrioventricular block, or digitalis intoxication. This study provides data pertaining to the safety of precordial pacing measured as the ratio of the minimum current required to induce ventricular fibrillation (VF) to the minimum current required to pace the ventricular myocardium. Single rectangular pulse stimuli having durations of 1-50 ms were evaluated. In 6 anesthetized dogs, a cutaneous electrode was centered over the shaved apex beat area of the left chest and paired with a larger electrode sutured to the right chest wall. Using a specially fabricated, synchronized, high-energy pulse generator, the "most vulnerable" time in the cardiac cycle at which the least current was required to induce VF with a single shock was identified. This current was compared to the much lower current required to pace the heart in the diastolic interval. While the current required to produce either fibrillation or pacing decreased as pulse duration increased, the safety factor (i.e., the ratio of fibrillation current to pacing current) remained nearly constant, averaging 12.6 for all pulse durations examined. That is, on the average, a stimulus of any given duration between 1 and 50 ms required at least 12 times the current required for pacing to produce ventricular fibrillation. We conclude that in normal canine hearts, the risk of inducing VF during precordial pacing is small.
No takes yet. Share an insight, caveat, or question.
Voorhees et al. (1984) studied Normal canine hearts (n=6). Precordial pacing was evaluated on Safety factor (ratio of minimum current required to induce ventricular fibrillation to minimum current required to pace). Precordial pacing in normal canine hearts demonstrated a high safety factor, requiring an average of 12.6 times the pacing current to induce ventricular fibrillation.
Synapse has enriched 10 related papers. Consider them for comparative context: