Key result
Valve design and leaflet motion significantly affected CT artifacts, with the Carbomedics valve and closed leaflet phases producing the least artifacts (p<0.001).
Why the study?
Does valve design and leaflet motion affect CT artifacts in mechanical prosthetic heart valves?
Does valve design and leaflet motion affect CT artifacts in mechanical prosthetic heart valves?
Valve design and leaflet motion significantly affect CT artifacts in mechanical prosthetic heart valves, with the best imaging results expected during phases when leaflets are closed.
May reduce misdiagnosis from artifacts in valve CT; leaves open human validation of design-specific protocols.
OBJECTIVES: Multidetector computed tomography (MDCT) can detect the cause of prosthetic heart valve (PHV) dysfunction but is hampered by valve-induced artifacts. We quantified artifacts of four PHV using a pulsatile in-vitro model and assessed the relation to leaflet motion and valve design. METHODS: A Medtronic Hall tilting disc (MH), and Carbomedics (CM), St Jude (SJM), and ON-X bileaflet valves underwent CT in an in-vitro model using retrospective gating with a 64 detector CT system in stationary and pulsatile conditions. Artifacts and radiopaque component volumes were quantified with thresholds based on surrounding structures and valvular components. RESULTS: Hypodense artifacts volumes (mm³) were 1,029 ± 147, 535 ± 53, 371 ± 16, and 366 ± 18 for the SJM, MH, CM and ON-X valves (p < 0.001 except for the latter two valves p = 0.43). Hyperdense artifact volumes were 3,546 ± 141, 2,387 ± 103, 2,003 ± 102, and 3,033 ± 31 for the SJM, MH, CM and ON-X valve, respectively (all differences p < 0.001). Leaflet motion affected hypodense (F = 41.5, p < 0.001) and hyperdense artifacts (F = 53.7, p < 0.001). Closed and moving leaflets were associated with the least and the most artifacts respectively (p < 0.001, both artifact types). CONCLUSION: Both valve design and leaflet motion affect PHV-induced artifacts. Best imaging results may be expected for the CM valve during phases in which the leaflets are closed.
No takes yet. Share an insight, caveat, or question.
Symersky et al. (2011) studied Prosthetic heart valve artifacts (n=4). Mechanical prosthetic heart valves (St Jude, Medtronic Hall, Carbomedics, ON-X) vs. Comparison between valve types and leaflet motion phases was evaluated on Hypodense and hyperdense artifact volumes (p=<0.001). Valve design and leaflet motion significantly affected CT artifacts, with the Carbomedics valve and closed leaflet phases producing the least artifacts (p<0.001).