Key result
Prospective ECG triggering significantly reduced hyperdense artefacts for both St Jude and Medtronic Hall prosthetic heart valves at 60, 75, and 90 beats per minute compared to retrospective gating.
Why the study?
Does prospective ECG triggering reduce prosthetic heart valve-induced artefacts compared with retrospective ECG gating on 256-slice CT in an in vitro model?
Does prospective ECG triggering reduce prosthetic heart valve-induced artefacts compared with retrospective ECG gating on 256-slice CT in an in vitro model?
Prospective ECG triggering on 256-slice CT significantly reduces hyper- and hypodense artefacts from mechanical prosthetic heart valves in vitro compared to retrospective gating, while using a lower radiation dose.
Answers are grounded in this paper and cite their sources.
May reduce prosthetic valve artefacts on CT in vitro; leaves open clinical translation.
OBJECTIVES: Multidetector computed tomography (MDCT) has diagnostic value for the evaluation of prosthetic heart valve (PHV) dysfunction but it is hampered by artefacts. We hypothesised that image acquisition using prospective triggering instead of retrospective gating would reduce artefacts related to pulsating PHV. METHODS: In a pulsatile in vitro model, a mono- and bileaflet PHV were imaged using 256 MDCT at 60, 75 and 90 beats per minute (BPM) with either retrospective gating (120 kV, 600 mAs, pitch 0.2, CTDI(vol) 39.8 mGy) or prospective triggering (120 kV, 200 mAs, CTDI(vol) 13.3 mGy). Two thresholds (>175 and <-45HU), derived from the density of surrounding structures, were used for quantification of hyper- and hypodense artefacts. Image noise and artefacts were compared between protocols. RESULTS: Prospective triggering reduced hyperdense artefacts for both valves at every BPM (P = 0.001 all comparisons). Hypodense artefacts were reduced for the monoleaflet valve at 60 (P = 0.009), 75 (P = 0.016) and 90 BPM (P = 0.001), and for the bileaflet valves at 60 (P = 0.001), 90 (P = 0.001) but not at 75 BPM (P = 0.6). Prospective triggering reduced image noise at 60 (P = 0.001) and 75 (P < 0.03) but not at 90 BPM. CONCLUSIONS: Compared with retrospective gating, prospective triggering reduced most artefacts related to pulsating PHV in vitro. KEY POINTS: • Computed tomographic images are often degraded by prosthetic heart valve-induced artefacts • Prospective triggering reduces prosthetic heart valve-induced artefacts in vitro • Artefact reduction at 90 beats per minute occurs without image noise reduction • Prospective triggering may improve CT image quality of moving hyperdense structures.
No takes yet. Share an insight, caveat, or question.
Symersky et al. (2011) studied Prosthetic heart valve-induced artefacts. Prospective ECG triggering vs. Retrospective ECG gating was evaluated on Hyperdense artefact volume (>175 HU) (p=<0.001). Prospective ECG triggering significantly reduced hyperdense artefacts for both St Jude and Medtronic Hall prosthetic heart valves at 60, 75, and 90 beats per minute compared to retrospective gating.