Key result
Multidetector CT provided higher image quality for evaluating bileaflet prosthetic heart valves compared to tilting disc valves, while Björk-Shiley and Sorin valves were non-diagnostic due to severe artifacts.
Why the study?
Does multidetector CT provide adequate image quality for the assessment of different types of prosthetic heart valves?
Observational (n=83)
Yes
Does multidetector CT provide adequate image quality for the assessment of different types of prosthetic heart valves?
Multidetector CT provides good image quality for most currently implanted prosthetic heart valves, but is non-diagnostic for older cobalt-containing tilting disc valves due to severe artifacts.
Answers are grounded in this paper and cite their sources.
CT image quality varies by valve type; supports targeted use in bileaflet prostheses but leaves open reliable assessment of older tilting-disc valves.
OBJECTIVES: Multidetector CT (MDCT) has shown potential for prosthetic heart valve (PHV) assessment. We assessed the image quality of different PHV types to determine which valves are suitable for MDCT evaluation. METHODS: All ECG-gated CTs performed in our institutions since 2003 were reviewed for the presence of PHVs. After reconstruction in 3 specific PHV planes, image quality of the supravalvular, perivalvular, subvalvular and valvular regions was scored on a four-point scale (1 = non-diagnostic, 2 = moderate, 3 = good and 4 = excellent) by two independent observers. RESULTS: Eighty-four CT examinations (66 cardiac, 18 limited-dose aortic protocols) of 83 patients with a total of 91 PHVs in the aortic (n = 71), mitral (n = 17), pulmonary (n = 1) and tricuspid (n = 2) position were included. CT was performed on a 16-slice (n = 4), 64-slice (n = 28) or 256-slice (n = 52) MDCT system. Median image quality scores for the supra-, peri- and subvalvular regions and valvular detail were (3.5, 3.3, 3.5 and 3.5, respectively) for bileaflet PHV; (3.0, 3.0, 3.5 and 3.0, respectively) for Medtronic Hall PHV; (1.0, 1.0, 1.0 and 1.0, respectively) for Björk-Shiley and Sorin monoleaflet PHV and (3.5, 3.5, 4.0 and 2.0 respectively) for biological PHV. CONCLUSION: Currently implanted PHVs have good image quality on MDCT and are suitable for MDCT evaluation.
No takes yet. Share an insight, caveat, or question.
Habets et al. (2011) conducted an observational in Prosthetic heart valves (n=83). Multidetector CT (MDCT) vs. Different PHV types (Bileaflet vs. Tilting disc) was evaluated on Median image quality score for valvular detail (1-4 scale) (p=<0.001). Multidetector CT provided higher image quality for evaluating bileaflet prosthetic heart valves compared to tilting disc valves, while Björk-Shiley and Sorin valves were non-diagnostic due to severe artifacts.