Key result
High-dose radiotherapy (47.25 Gy) applied in close proximity to a cardiac resynchronization therapy device was carried out safely without device malfunction over >1.5 years of follow-up.
Why the study?
Guidelines and manufacturer instructions lack practical recommendations for cumulative radiation doses to pacemakers or implantable cardioverter defibrillators in patients undergoing radiotherapy.
Does high-dose radiotherapy applied in close proximity to a cardiac resynchronization therapy device cause device malfunction?
Case Report (n=1)
Does high-dose radiotherapy applied in close proximity to a cardiac resynchronization therapy device cause device malfunction?
High-dose radiotherapy near a cardiac resynchronization therapy device can be performed safely without device relocation, provided close monitoring is maintained.
Patients with cardiac implantable electronic devices undergoing radiotherapy (RT) are prone to the risk of device failure. Guidelines and manufacturer's instructions are lacking practical recommendations for cumulative radiation doses to pacemakers or implantable cardioverter defibrillators. The present case demonstrates the effect of RT of a Merkel cell carcinoma near the location of a cardiac resynchronization therapy pacemaker. Despite guideline recommendations, surgical relocation or de novo implantation of the device on the contralateral side was avoided to prevent the dissemination of tumour cells, inflammation, and wound healing complications. A total dose of 47.25 Gy applied in very close proximity to the cardiac resynchronization therapy pacemaker was carried out safely without jeopardizing the patient and any device malfunction during and after treatment within >1.5 years of follow-up period. The present case demonstrates that high-dose RT near to a cardiac resynchronization therapy device can be carried out safely. Special precautions during RT as well as close device follow-up interrogations are mandatory. Large-scale studies are needed for the true frequency of adverse events.
No takes yet. Share an insight, caveat, or question.
Schernthaner et al. (2020) conducted a case report in Merkel cell carcinoma near a cardiac resynchronization therapy pacemaker (n=1). Radiotherapy was evaluated on Device malfunction. High-dose radiotherapy (47.25 Gy) applied in close proximity to a cardiac resynchronization therapy device was carried out safely without device malfunction over >1.5 years of follow-up.