Key result
Radiotherapy with standardized protocols causes zero device malfunctions or major adverse events in CIED carriers.
Why the study?
The number of patients with cardiac implantable electronic devices receiving radiotherapy is increasing, and their management is challenging and requires a collaborative multidisciplinary approach.
What is the incidence of device malfunction and adverse clinical events in patients with MRI-conditional cardiac implantable electronic devices undergoing radiotherapy under a standardized periprocedural reprogramming protocol?
Cohort (n=107)
No
What is the incidence of device malfunction and adverse clinical events in patients with MRI-conditional cardiac implantable electronic devices undergoing radiotherapy under a standardized periprocedural reprogramming protocol?
Answers are grounded in this paper and cite their sources.
This 10-year single-center experience demonstrates that radiotherapy can be safely delivered to patients with MRI-conditional pacemakers and ICDs without causing device malfunctions or clinically significant adverse events when managed with a standardized periprocedural reprogramming protocol.
BACKGROUND: The number of patients with cardiac implantable electronic devices (CIEDs) receiving radiotherapy (RT) is increasing. The management of CIED-carriers undergoing RT is challenging and requires a collaborative multidisciplinary approach. AIM: The aim of the study is to report the real-world, ten-year experience of a tertiary multidisciplinary teaching hospital. METHODS: We conducted an observational, real-world, retrospective, single-center study, enrolling all CIED-carriers who underwent RT at the San Raffaele University Hospital, between June 2010 and December 2021. All devices were MRI-conditional. The devices were programmed to an asynchronous pacing mode for patients who had an intrinsic heart rate of less than 40 beats per minute. An inhibited pacing mode was used for all other patients. All tachyarrhythmia device functions were temporarily disabled. After each RT session, the CIED were reprogrammed to the original settings. Outcomes included adverse events and changes in the variables that indicate lead and device functions. RESULTS: Between June 2010 and December 2021, 107 patients were enrolled, among which 63 (58.9%) were pacemaker carriers and 44 (41.1%) were ICD carriers. Patients were subjected to a mean of 16.4 (±10.7) RT sessions. The most represented tumors in our cohort were prostate cancer (12; 11%), breast cancer (10; 9%) and lung cancer (28; 26%). No statistically significant changes in device parameters were recorded before and after radiotherapy. Generator failures, power-on resets, changes in pacing threshold or sensing requiring system revision or programming changes, battery depletions, pacing inhibitions and inappropriate therapies did not occur in our cohort of patients during a ten-year time span period. Atrial arrhythmias were recorded during RT session in 14 patients (13.1%) and ventricular arrhythmias were observed at device interrogation in 10 patients (9.9%). CONCLUSIONS: Changes in device parameters and arrhythmia occurrence were infrequent, and none resulted in a clinically significant adverse event.
No takes yet. Share an insight, caveat, or question.
Gulletta et al. (2022) conducted a cohort in Cancer patients with MRI-conditional cardiac implantable electronic devices (CIEDs) undergoing radiation therapy (n=107). Radiotherapy with standardized periprocedural CIED reprogramming (asynchronous pacing if intrinsic HR <40 bpm, inhibited pacing otherwise, and temporary disabling of tachyarrhythmia functions) was evaluated on Device malfunctions and major adverse events (generator failure, power-on reset, changes in pacing threshold or sensing requiring revision or programming changes, battery depletion, pacing inhibition, inappropriate therapies). Radiotherapy in 107 MRI-conditional CIED carriers resulted in 0% device malfunctions or major adverse events and no significant changes in lead or device parameters.