Key result
During uterine intracavitary brachytherapy with a total prescribed dose of 30 Gy, the estimated dose at the surface of the ICD position was 21.2 mGy, well below the 1 Gy maximum recommended limit.
Why the study?
Does uterine intracavitary brachytherapy using an 192Ir sealed source deliver a safe radiation dose to an implanted ICD?
Does uterine intracavitary brachytherapy using an 192Ir sealed source deliver a safe radiation dose to an implanted ICD?
Uterine intracavitary brachytherapy using an 192Ir sealed source delivers a radiation dose to an implanted ICD well below the recommended 1 Gy safety limit, suggesting it can be safely performed with appropriate monitoring.
Answers are grounded in this paper and cite their sources.
Hypothesis-generating for safe brachytherapy in ICD patients; prospective studies needed before practice change.
The purpose of this study was to verify the dose absorbed by an implantable cardioverter defibrillator (ICD) from an (192)Ir sealed source during uterine intracavitary brachytherapy, and to confirm its immunity to radiation effects. First, prior to treatment, the doses around the ICD position of an anthromorphic phantom were evaluated. Next, we also measured the dose at the ICD position using a fluorescent glass dosimeter and silicon diode dosimeter during the treatment of intracavitary brachytherapy of a patient implanted with an ICD. The results of the phantom study showed the dose percentage at the ICD location, 2 cm deep, to be 0.074% of the prescribed dose. The results of a treatment study similarly showed the dose, measured using a fluorescent glass dosimeter in the ICD position, to be 0.071% of the prescribed dose. During the application of the total prescribed dose, 30 Gy/5 fraction, the dose at the surface of the ICD position was estimated to be 21.2 mGy, well below the 1 Gy maximum recommended in the JASTRO guidelines. We regard dose verification and monitoring during treatment to be both necessary and useful in the treatment of individual cases.
No takes yet. Share an insight, caveat, or question.
Hirose et al. (2014) studied Uterine cancer requiring intracavitary brachytherapy in a patient with an ICD (n=1). Uterine intracavitary brachytherapy (192Ir sealed source) was evaluated on Radiation dose absorbed at the ICD position. During uterine intracavitary brachytherapy with a total prescribed dose of 30 Gy, the estimated dose at the surface of the ICD position was 21.2 mGy, well below the 1 Gy maximum recommended limit.
Synapse has enriched 10 related papers. Consider them for comparative context: