Key result
A 3D circuit model-based Bio-Impedance simulation platform was developed and validated by experimental measurements to optimize arterial pulse wave monitoring prior to experimentation.
Why the study?
Bio-Impedance measured data for arterial pulse wave monitoring are significantly affected by electrode placement relative to the artery and electrode configuration.
A novel 3D circuit model simulation platform for Bio-Impedance sensors can optimize electrode placement and configuration for continuous arterial pulse wave monitoring prior to physical experimentation.
Answers are grounded in this paper and cite their sources.
Preclinical simulation platform aids electrode optimization for bio-impedance monitoring; leaves open clinical translation pending human validation.
Cardiovascular disease (CVD) threatens the lives of many and affects their productivity. Wearable sensors can enable continuous monitoring of hemodynamic parameters to improve the diagnosis and management of CVD. Bio-Impedance (Bio-Z) is an effective non-invasive sensor for arterial pulse wave monitoring based on blood volume changes in the artery due to the deep penetration of its current signal inside the tissue. However, the measured data are significantly affected by the placement of electrodes relative to the artery and the electrode configuration. In this work, we created a Bio-Z simulation platform that models the tissue, arterial pulse wave, and Bio-Z sensing configuration using a 3D circuit model based on a time-varying impedance grid. A new method is proposed to accurately simulate the different tissue types such as blood, fat, muscles, and bones in a 3D circuit model in addition to the pulsatile activity of the arteries through a variable impedance model. This circuit model is simulated in SPICE and can be used to guide design decisions (i.e. electrode placement relative to the artery and electrode configuration) to optimize the monitoring of pulse wave prior to experimentation. We present extensive simulations of the arterial pulse waveform for different sensor locations, electrode sizes, current injection frequencies, and artery depths. These simulations are validated by experimental Bio-Z measurements.
No takes yet. Share an insight, caveat, or question.
Ibrahim et al. (2021) studied Cardiovascular disease. Bio-Impedance (Bio-Z) simulation platform using a 3D circuit model was evaluated on Arterial pulse waveform simulation and validation. A 3D circuit model-based Bio-Impedance simulation platform was developed and validated by experimental measurements to optimize arterial pulse wave monitoring prior to experimentation.
Synapse has enriched 10 related papers. Consider them for comparative context: