Answers are grounded in this paper and cite their sources.
PURPOSE: To describe an in vitro feasibility trial of a new percutaneous endograft delivery technique. METHODS AND RESULTS: A water flow model of 9-mm (inner diameter) transparent plastic tubing was used to test the feasibility of sequentially delivering the components of an endograft through a 7-F sheath for assembly in situ. The tubular endovascular graft was fabricated from a 10-mm x 68-mm Wallstent and 50-microm-thick Dacron graft. The graft material was formed into a tube, attached with a suture to a guidewire, and delivered into the plastic tubing. The Wallstent was then delivered through the tubular graft and deployed, affixing the graft to the plastic tube wall. In 4 trials, only 1 attempt was not successful. CONCLUSIONS: These concepts and techniques may have implications in the development of percutaneously deliverable endovascular grafts.
No takes yet. Share an insight, caveat, or question.
Kerr et al. (2000) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: