Answers are grounded in this paper and cite their sources.
A low temperature thermometer based on the temperature dependence of the vF(T) of the 19F NMR frequency in antiferromagnetic MnF2 has been successfully operated in the 10–40 K temperature region. Novel features include the use of a MnF2 crystal with a sharp 19F NMR line as the temperature sensitive element in a solid state frequency-locking autodyne circuit that operates while completely submerged in liquid H2 or He. In the present design the circuit-sample combination takes the form of a compact probe. With this device temperature readings which are reproducible to ±10−4 K at 20 K have been obtained. Various applications for precision thermometry in the 10–40 K range using the device are proposed.
No takes yet. Share an insight, caveat, or question.
Gill et al. (1969) studied this question.