Key result
Atrial (AAI) pacing in patients with sick sinus syndrome was safe over a mean 25 months, with 20.9% (9/43) experiencing mild deterioration in atrioventricular conduction.
Why the study?
Is atrial (AAI) pacing safe and effective in patients with sick sinus syndrome?
Cohort (n=43)
Is atrial (AAI) pacing safe and effective in patients with sick sinus syndrome?
Atrial (AAI) pacing is a safe and reliable option for patients with sick sinus syndrome, with a low risk of subsequent symptomatic conduction system disease.
Supports short-term AAI pacing safety in sick sinus syndrome with monitoring; leaves open randomized long-term AV conduction comparisons.
Atrial pacing has the advantages of simplicity, maintenance of AV synchrony, and economy. The major detraction has been the potential for deterioration of atrioventricular conduction. In this study, we followed 43 patients with sick sinus syndrome treated with atrial (AAI) pacing. Excellent initial implant parameters were obtained in all. Three early lead repositionings were required. Minor sensing and pacing problems could all otherwise be handled by reprogramming. Follow-up for a mean of 25 +/- 20 months demonstrated excellent performance of the pacing systems. Pacing and sensing thresholds and lead impedance indicated excellent lead performance. There were no late lead failures. Nine patients have had mild deterioration in atrioventricular conduction and one of these had a change to DDD pacing at the time of elective battery change. All patients are asymptomatic. Thus, chronic atrial pacing in selected patients is safe and reliable with good chronic lead performance and low risk of subsequent conduction system disease.
No takes yet. Share an insight, caveat, or question.
Kerr et al. (1989) conducted a cohort in Sick sinus syndrome (n=43). Atrial (AAI) pacing was evaluated on Mild deterioration in atrioventricular conduction. Atrial (AAI) pacing in patients with sick sinus syndrome was safe over a mean 25 months, with 20.9% (9/43) experiencing mild deterioration in atrioventricular conduction.