Key result
Platinized groove surface electrodes had lower chronic capture thresholds at 10 months compared to ring-tip electrodes, though clinically inconsequential.
Why the study?
Does a platinized groove surface electrode improve acute and chronic capture thresholds compared to a ring-tip electrode in patients receiving a dual chamber pacemaker?
Observational (n=47)
Does a platinized groove surface electrode improve acute and chronic capture thresholds compared to a ring-tip electrode in patients receiving a dual chamber pacemaker?
Answers are grounded in this paper and cite their sources.
Platinized grooved electrodes provide significantly lower chronic capture thresholds than ring-tip electrodes, though the difference is small and likely clinically inconsequential.
One of the determinants of the capture threshold of an endocardial pacing lead is the configuration of the electrode tip. To evaluate whether micro- and macroporous electrodes have better initial and chronic thresholds than nonporous electrodes, acute and chronic capture thresholds, stimulation impedance and sensing thresholds were determined in 22 patients in whom a ventricular ring-tip electrode and a unipolar, dual chamber pacemaker with bidirectional telemetry had been implanted. These values were compared to those obtained from 25 patients receiving an electrode constructed with a platinized groove surface at the time of implant of an identical pulse generator. The ventricular capture threshold at implant was 0.7 V +/- 0.3 at 0.6 msec pulse width for both groups. The capture threshold was significantly greater in the ring tip electrode group at follow-up periods of 1 month (1.1 V +/- 0.5 vs 1.6 V +/- 0.6, P less than 0.008), 4 months (1.0 V +/- 0.2 vs 1.7 V +/- 0.8, P less than 0.002), and 10 months (1.2 V +/- 0.4 vs 1.7 V +/- 0.5, P less than 0.04) following implantation. The stimulation impedance at the time of implantation was lower in the ring-tip electrode group (530 ohms vs 603 ohms, P less than 0.03), but thereafter no significant difference was seen between the two groups. The acute and chronic sensing thresholds were similar in both groups. While the microporous electrode had significantly lower chronic capture thresholds, the magnitude of this difference is small, and probably clinically inconsequential.(ABSTRACT TRUNCATED AT 250 WORDS)
No takes yet. Share an insight, caveat, or question.
Midei et al. (1989) conducted an observational in Patients requiring a unipolar, dual chamber pacemaker (n=47). Platinized groove surface electrode vs. Ventricular ring-tip electrode was evaluated on Capture threshold at 10 months (p=<0.04). Platinized groove surface electrodes had lower chronic capture thresholds at 10 months compared to ring-tip electrodes, though clinically inconsequential.
Synapse has enriched 10 related papers. Consider them for comparative context: