Why the study?
Does a carbon-tip endocardial electrode reduce chronic stimulation threshold compared to a platinum-tip electrode in patients receiving VVI pacemakers?
Does a carbon-tip endocardial electrode reduce chronic stimulation threshold compared to a platinum-tip electrode in patients receiving VVI pacemakers?
Carbon-tip endocardial electrodes provide a lower chronic stimulation threshold at 15 months compared to platinum-tip electrodes, allowing for reduced pacemaker output in more patients.
Answers are grounded in this paper and cite their sources.
Carbon-tip electrodes may enable lower outputs in VVI pacemakers; extends material comparisons for chronic threshold reduction.
Seventy-nine patients admitted for VVI pacemaker treatment were randomly given a Siemens-Elema 412 S endocardial electrode (tip material vitreous carbon, tip area 12 mm2) (41 patients) or a Siemens-Elema 415 S endocardial electrode (tip material platinum, tip area 12 mm2) (38 patients). The groups did not differ in age, sex, magnitude of intracardiac signal or indication for pacemaker treatment. Pacing threshold was measured by the vario technique at 3 and 15 months after the implantation. In the group of carbon-tip electrodes and in the group of platinum-tip electrodes stimulation thresholds at implantation were 0.49 +/- 0.10 V and 0.54 +/- 0.14 V (ns), at months 31.40 +/- 0.45 V and 1.57 +/- 0.53 V (ns), and at months 15 1.12 +/- 0.33 V and 1.37 +/- 0.38 V (P = 0.003), respectively. After 1 year, 88% of the patients in the carbon-tip group and 66% (P = 0.04) in the platinum-tip group fulfilled our criteria for reducing pacemaker output from 5.0 V to 2.5 V. In comparable platinum-tip and carbon-tip pacing electrodes, the chronic stimulation threshold was lower in the carbon-tip than in the platinum-tip electrode.
No takes yet. Share an insight, caveat, or question.
Thuesen et al. (1989) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: