Key result
Suboptimal THV implantation linked to ~3-fold higher odds of early safety events vs. optimal placement.
Why the study?
The expansion of TAVR to younger patients with longer life expectancy requires procedural optimization to enhance long-term valve durability, prompting evaluation of how fluoroscopic determinants of optimal implantation relate to clinical and valve outcomes.
Does optimal THV implantation improve early procedural safety and long-term THV durability in patients undergoing balloon-expandable TAVR?
Cohort (n=1,032)
No
Does optimal THV implantation improve early procedural safety and long-term THV durability in patients undergoing balloon-expandable TAVR?
Answers are grounded in this paper and cite their sources.
Optimal THV implantation, defined by favorable coaxiality, depth, and expansion, is associated with improved early procedural safety and reduced long-term bioprosthetic valve failure in balloon-expandable TAVR.
BACKGROUND: The expansion of transcatheter aortic valve replacement (TAVR) to younger patients with longer life expectancy requires procedural optimization to enhance long-term valve durability. OBJECTIVES: To evaluate the relationship between fluoroscopic determinants of optimal transcatheter heart valve (THV) implantation-coaxiality, implantation depth, and expansion-and clinical and valve-related outcomes. METHODS: In this cohort study, consecutive patients undergoing transfemoral TAVR with balloon-expandable prostheses (Sapien S3, Sapien S3 Ultra) between February 2014 and June 2022 were included. THV coaxiality, implantation depth, and expansion were assessed in the 3-cusp view following deployment using standardized protocols. Suboptimal implantation was defined as ≥1 unfavorable determinants (coaxiality ≥4.8°, depth ≥6 mm, or underexpansion ≥20%). Outcomes included the modified early safety endpoint, moderate or severe hemodynamic valve deterioration, and stage 2 bioprosthetic valve failure. RESULTS: Among 1,032 patients (mean age 81 ± 6 years, 35.1% female), suboptimal THV implantation occurred in 443 (42.9%) patients and was driven by noncoaxial deployment (n = 240, 23.3%), excessive implantation depth (n = 261, 25.3%), and underexpansion (n = 54, 5.2%). Suboptimal implantation was associated with a higher likelihood of the modified early safety endpoint at 30 days (43.3% vs 19.9%; adjusted odds ratio: 3.05; 95% CI: 2.30-4.04) and stage 2 bioprosthetic valve failure at 5 years (subHR: 4.40; 95% CI: 1.25-13.59). Patients with ≥2 determinants demonstrated higher rates of hemodynamic valve deterioration at 5 years (subHR: 3.48; 95% CI: 1.24-9.80). CONCLUSIONS: Optimal THV implantation was associated with improved early procedural safety and long-term THV durability in balloon-expandable TAVR. Systematic fluoroscopic assessment of optimized valve implantation may enhance patient outcomes.
No takes yet. Share an insight, caveat, or question.
Vitez et al. (2026) conducted a cohort in Severe aortic stenosis (n=1,032). Suboptimal THV implantation vs. Optimal THV implantation was evaluated on Modified early safety endpoint at 30 days (aOR 3.05, 95% CI 2.30-4.04, p=<0.001). Suboptimal transcatheter heart valve implantation was associated with a higher likelihood of the modified early safety endpoint at 30 days compared to optimal implantation.