Key result
Higher balloon-expandable THV asymmetry links to ~37% higher risk of impaired hemodynamic performance.
Why the study?
Asymmetrical expansion of transcatheter heart valves occasionally occurs during TAVR, but its impact on hemodynamic valve performance and clinical outcomes was unknown.
Does asymmetrical expansion of balloon-expandable THVs predict impaired hemodynamic valve performance and clinical outcomes in patients undergoing TAVR?
Observational (n=1,216)
Does asymmetrical expansion of balloon-expandable THVs predict impaired hemodynamic valve performance and clinical outcomes in patients undergoing TAVR?
Answers are grounded in this paper and cite their sources.
Captured external expert commentary on this paper, strongest first. Original sources are linked where available.
“We found that asymmetrical expansion of the THV was associated with hemodynamic valve performance. Patients with high degrees of prosthetic valve asymmetry had a higher risk for increased residual transvalvular gradients … moreover, our analysis suggests that patients with valve asymmetry have a higher risk for relevant paravalvular aortic regurgitation. Hemodynamic valve performance may be a surrogate marker for long-term clinical outcomes.”
Asymmetrical expansion of balloon-expandable THVs during TAVR strongly predicts impaired hemodynamic valve performance but does not appear to affect mid-term all-cause mortality.
BACKGROUND: Asymmetrical expansion of transcatheter heart valves (THVs), manifesting as stent frame deformation, is an occasional fluoroscopic finding in transcatheter aortic valve replacement (TAVR). OBJECTIVES: The aim of this study was to investigate the impact of asymmetrical expansion of balloon-expandable THVs on hemodynamic valve performance and clinical outcomes. METHODS: In a prospective registry, TAVR asymmetry index was measured using freeze-frame fluoroscopic images and was defined as the ratio of THV heights: [(longer height/shorter height) - 1] × 100. THV hemodynamic performance was measured using echocardiography before hospital discharge. Impaired hemodynamic valve performance was defined as a mean residual THV gradient ≥20 mm Hg and/or moderate or greater paravalvular regurgitation. RESULTS: Among 1,216 patients undergoing transfemoral TAVR for native severe aortic valve stenosis with contemporary balloon-expandable THVs between February 2014 and June 2022, asymmetry index was an excellent predictor of impaired hemodynamic valve performance (area under the receiver-operating characteristic curve: 0.88; 95% CI: 0.84-0.92; P < 0.001). The optimal asymmetry index threshold for predicting impaired THV performance was >5.5% (sensitivity 77%, specificity 86%) and occurred in 17% of patients. Higher asymmetry index, as a continuous variable, was associated with impaired hemodynamic valve performance independent of total aortic valve calcium, bicuspid anatomy, balloon-expandable valve prosthesis type or size, and Society of Thoracic Surgeons Predicted Risk of Mortality score (OR: 1.37; 95% CI: 1.29-1.46; P < 0.001). High TAVR asymmetry index was not associated with all-cause mortality during a median of 376 days of follow-up (HR: 1.00; 95% CI: 0.71-1.41; P = 0.989). CONCLUSIONS: Asymmetrical expansion of balloon-expandable THVs was associated with impaired hemodynamic valve performance but not with clinical outcomes.
No takes yet. Share an insight, caveat, or question.
Maznyczka et al. (2024) conducted an observational in Native severe aortic valve stenosis (n=1,216). Higher asymmetry index of balloon-expandable THVs vs. Lower asymmetry index was evaluated on Impaired hemodynamic valve performance (mean residual THV gradient ≥20 mm Hg and/or moderate or greater paravalvular regurgitation) (OR 1.37, 95% CI 1.29-1.46, p=<0.001). Higher asymmetry index of balloon-expandable transcatheter heart valves was associated with impaired hemodynamic valve performance but not all-cause mortality.
Synapse has enriched 10 related papers. Consider them for comparative context: