Key result
In patients with bicuspid aortic stenosis, geometric deformation of self-expanding THVs at the inflow did not compromise immediate valve function, hemodynamics, or leaflet integrity.
Why the study?
Does geometric deformation of self-expanding transcatheter heart valves affect valve performance in patients with bicuspid aortic stenosis?
Does geometric deformation of self-expanding transcatheter heart valves affect valve performance in patients with bicuspid aortic stenosis?
Answers are grounded in this paper and cite their sources.
In bicuspid aortic stenosis, geometric deformation of self-expanding transcatheter heart valves at the inflow does not compromise immediate valve function, though calcified raphe is associated with increased paravalvular leak.
This editorial refers to ‘Stent frame deformation of self-expanding transcatheter heart valves in bicuspid aortic stenosis and impact on valve performance’, by G.T. Conte et al., https://doi.org/10.1093/ehjci/jeaf195. Knowledge is a process of piling up facts; wisdom lies in their simplification. (Martin Fischer) Transcatheter aortic valve implantation (TAVI) in patients with bicuspid aortic valves (BAVs) presents unique technical challenges due to the complex and heterogeneous anatomy characteristic of this population. Variations in leaflet morphology, the presence of eccentric and heavily calcified raphe, and frequent dilation of the ascending aorta complicate procedural planning and device deployment.1 Self-expanding, supra-annular transcatheter heart valves (THVs) are often favoured in this setting, as they are associated with a reduced risk of annular rupture compared with balloon-expandable (BE) platforms. Moreover, supra-annular THVs typically achieve larger effective orifice areas, which may translate into superior haemodynamic performance. Nonetheless, the substantial calcific burden seen in BAVs frequently raises concerns regarding incomplete or elliptical stent frame expansion and the potential implications for valve haemodynamics, durability, and long-term clinical outcomes. In this issue, Conte et al.2 report the findings of a multicenter registry study evaluating stent frame expansion and ellipticity of the self-expanding Evolut™ THV (Medtronic) in patients with bicuspid aortic stenosis (AS). Using pre- and post-procedural computed tomography (CT), the authors analysed 175 cases to assess the relationship between stent geometry and valve performance. Valve performance was evaluated through mean transvalvular gradient (mmHg), the presence and severity of paravalvular leak (PVL), and hypoattenuated leaflet thickening, a surrogate marker for leaflet thrombosis. The study demonstrated greater under-expansion and ellipticity at the inflow level (LVOT) compared with the leaflet level of the Evolut™ THV. Specifically, the ellipticity of the stent was significantly higher at the LVOT, then the leaflet level. Similarly, the expansion of the stent was nearly 95% compared with about 75% expansion at the LVOT level. Importantly, neither inflow under-expansion nor ellipticity adversely affected transvalvular haemodynamics, PVL, or leaflet thickening. Interestingly, the study by Conte et al. found the mean post-TAVI transvalvular gradient to be 8.2 ± 3.7 mmHg in their cohort of patients with BAV. This compared well with the 1-year post-TAVI gradients in the Evolut™ low-risk trial (8.6 ± 3.7 mmHg).3 These findings are important as prior studies have suggested that the unique anatomy of bicuspid valves may predispose patients to incomplete THV expansion, with associated risks of PVL, leaflet thrombosis, and concerns regarding long-term durability.3 As such, this study provides reassuring evidence that despite geometric deformation at the inflow of the SE-THV, valve function and performance were preserved. The study further evaluated anatomical factors influencing stent frame geometry and valve performance within this BAV cohort. Notably, aortic annulus size was significantly associated with the degree of stent frame expansion at the leaflet level; however, this variation in expansion did not translate into differences in Doppler transvalvular mean gradient, indicating preserved haemodynamic performance irrespective of annular dimensions. Additionally, Sievers type 1 BAV morphology was associated with a greater degree of inflow under-expansion when compared with type 0 anatomy. The presence of a calcified raphe correlated with increased inflow ellipticity, underscoring the role of asymmetric calcification in influencing stent geometry. Despite these anatomic challenges, stent frame circularity and expansion at the leaflet level were maintained, and no adverse effects on THV haemodynamics or function were observed. However, it is important to note that long-term follow-up still needs to be conducted for this group of patients. The median duration of the follow-up was about 2 months post-TAVI. It is also important to note that the study found a significant association between calcified raphe and moderate-to-severe PVL (10.8% vs. 1.9%). This is the balance that clinicians must strike when choosing a THV platform in patients with BAVs. Over inflation of a balloon and the risk of annular tissue injury, or areas of geometric deformation that can cause PVL. Determining this balance in a patient-specific manner can be aided through computational modelling and the simulation of both BE and SE THV deployment. Figure 1 illustrates how pre-procedural aortic root CT may be used to plan TAVI in BAV and predict PVL post-TAVI. Also, although post-TAVI transvalvular gradients were promising in the study by De Backer et al., they do mention an important limitation regarding the lack of long-term follow-up in this cohort. Additionally, under-expansion and ellipticity of the index THV did not seem to impact haemodynamics and valve performance, but it may have important implications for future TAV-in-TAV procedures. This is even more relevant to the bicuspid AS patient population, who are often younger at the time of their index valve implant compared with patients with tricuspid aortic valve anatomy. Furthermore, deformation at the inflow of the secondary Evolut™, placed inside an index THV (BE or SE), may cause leaflet flutter, which has the potential to cause long-term concerns with valve leaflet durability due to an increase in turbulent blood flow.4 Computational modelling can assist in lifetime management planning for younger bicuspid patients to ensure that index valve implants optimize the ability for subsequent future transcatheter interventions (Figure 2). (A and B) 3D digital reconstruction of the pre-deployment native anatomy of a patient with Sievers type 1 BAV and a patient with Sievers type 0 BAV. Simulations using the FDA-cleared PrecisionTAVI software platform (DASI Simulations, Columbus, OH, USA) demonstrate the deformation of an Evolut™ THV in both patients (C and D). From the deformation information, a reduced-order fluids model can predict the most likely radial blood flow gaps, as well as measure their size (E and F). This may allow clinicians to understand how patient-specific stent deformation will determine PVL gap size and location. 3D, three-dimensional; BAV, bicuspid aortic valve; THV, transcatheter heart valve; PVL, paravalvular leak. (A and B) TAV-in-TAV simulation results with respect to the left coronary arteries of two patients using a simulated BE 26 THV in a BE 26 THV. (C and D) The same analysis for the right coronary arteries of the same two patients. (E and F) TAV-in-TAV simulation results with respect to the left coronary arteries of two patients using a simulated SE 29 in a BE 26. (G and H) The same analysis for the right coronary arteries of the same two patients. TAV-in-TAV, transcatheter aortic valve in transcatheter aortic valve; BE, balloon-expandable; THV, transcatheter heart valve; SE, self-expandable. The study by Conte et al. offers valuable evidence that, despite the complex anatomy inherent to BAVs, geometric deformation of self-expanding THVs—particularly at the inflow—does not compromise immediate valve function, haemodynamics, or leaflet integrity. These findings challenge prior concerns regarding the impact of stent frame under-expansion and ellipticity on valve performance and durability. However, the association between the calcified raphe and moderate-to-severe PVL highlights the ongoing need for careful procedural planning and individualized device selection. Computational modelling and patient-specific simulations may play an increasingly important role in guiding these decisions and optimizing outcomes in bicuspid AS patients undergoing TAVI: machines have the answers, but humans hold the questions. These are the questions that drive progress, and the real magic happens when we harmonize these forces. Data will be made available on request.
No takes yet. Share an insight, caveat, or question.
Becker et al. (2025) conducted an editorial in bicuspid aortic stenosis (n=175). Self-expanding Evolut THV was evaluated on Stent frame expansion and ellipticity and their relationship with valve performance. In patients with bicuspid aortic stenosis, geometric deformation of self-expanding THVs at the inflow did not compromise immediate valve function, hemodynamics, or leaflet integrity.