Key result
Evolut THV inflow stent expansion averages ~80% without compromising short-term valve hemodynamics.
Why the study?
In TAVI for bicuspid AS, concerns exist regarding reduced stent frame expansion and increased ellipticity, and the implications of these deformations on THV performance remain unclear.
Does stent frame deformation of the self-expanding Evolut THV impact valve performance in patients with bicuspid aortic stenosis?
Observational (n=175)
Yes
Does stent frame deformation of the self-expanding Evolut THV impact valve performance in patients with bicuspid aortic stenosis?
In patients with bicuspid aortic stenosis undergoing TAVI, stent frame under-expansion and ellipticity at the inflow level of the Evolut THV do not compromise short-term valve hemodynamics or cause leaflet thickening.
Answers are grounded in this paper and cite their sources.
Inflow under-expansion of Evolut THV was not associated with impaired short-term hemodynamics in bicuspid TAVI; leaves open questions on durability.
AIMS: To evaluate stent frame expansion and ellipticity of the Evolut™ transcatheter heart valve (THV) and its subsequent impact on valve performance in patients with bicuspid aortic stenosis (AS) using pre- and post-procedural cardiac computed tomography (CT). In transcatheter aortic valve implantation (TAVI) for native bicuspid AS, concerns arise regarding the risk for reduced stent frame expansion and increased ellipticity compared with valves implanted in tricuspid AS. The implications of stent frame under-expansion and eccentricity on THV performance remain unclear. METHODS AND RESULTS: This multi-centre registry included patients with bicuspid AS who underwent TAVI using the self-expanding Evolut™ THV and who had pre- and post-TAVI CT. Stent frame expansion and ellipticity were assessed at the inflow (nodes 0-1) and leaflet level (nodes 5-6). THV performance was evaluated by assessing mean transprosthetic gradient, paravalvular leak, and hypoattenuated leaflet thickening. Among 175 patients included, the inflow level of the Evolut THV had greater stent frame under-expansion (79.9% ± 6.7% vs. 95.4% ± 3.1% and P < 0.001) and ellipticity (1.3 ± 0.3 vs. 1.2 ± 0.1 and P < 0.001) compared to the leaflet level. Leaflet level expansion was unaffected by expansion or ellipticity at the inflow level; however, leaflet level ellipticity increased significantly with inflow under-expansion <75% (P = 0.002) and inflow ellipticity ≥1.3 (P < 0.001). Neither inflow under-expansion nor ellipticity impacted hemodynamic valve performance or leaflet thickening at short-term follow-up. CONCLUSION: The self-expanding Evolut™ THV demonstrates consistent valve performance in bicuspid AS. Stent frame under-expansion and ellipticity at the inflow level do not compromise valve haemodynamics or contribute to leaflet thickening at short-term follow-up.
No takes yet. Share an insight, caveat, or question.
Tirado‐Conte et al. (2025) conducted an observational in bicuspid aortic stenosis (n=175). Evolut transcatheter heart valve was evaluated on Stent frame expansion at the inflow level compared to the leaflet level (p=<0.001). The Evolut THV had greater stent frame under-expansion at the inflow level than the leaflet level, but this did not compromise short-term valve hemodynamics.