Key result
Implantation of a new soft total artificial heart with bistable polyurethane valves in calves resulted in low numbers of thromboemboli compared with previous experiments.
New bistable polyurethane valves in a soft total artificial heart show promising hemodynamics and low thrombogenicity in animal models.
Does not yet support clinical use; extends preclinical evidence for bistable valves in total artificial hearts.
This article describes new bistable valves, and introduces a new soft heart that is easy to implant. Earlier, five polyurethane (PU) valves were implanted in the mitral position in sheep. All five survived for 1 year or more, and the valves, although calcified, were intact. Since the opening resistance was somewhat high, valves that are bistable were developed, which means they may be open or closed. These valves have lower opening resistance, and regurgitation is similar to that of mechanical valves. Eight calves have been implanted with a new, soft total artificial heart (TAH). Seven had bistable leaflet valves; the eighth had mechanical (Bicer) valves in the inflow position. Four of the calves were sacrificed after 22 to 43 days. At autopsy, the number of thromboemboli found, particularly in the kidneys, was low compared with previous experiments. None of these animals received anticoagulants other than the heparin given during heart/lung bypass.
No takes yet. Share an insight, caveat, or question.
Yu et al. (1989) studied Total artificial heart implantation (n=13). Polyurethane bistable valves and soft total artificial heart vs. Mechanical valves and previous experiments was evaluated on Survival and thromboemboli at autopsy. Implantation of a new soft total artificial heart with bistable polyurethane valves in calves resulted in low numbers of thromboemboli compared with previous experiments.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: