Key result
Non-invasive cardiac output measurement using the T-Line system showed a mean difference of -0.8 l/minute (95% limits of agreement -4.0 to 2.3) compared to transpulmonary thermodilution.
Why the study?
Does continuous non-invasive cardiac output monitoring using radial artery applanation tonometry agree with invasive thermodilution and pulse contour analysis in ICU patients after major gastrointestinal surgery?
Observational (n=27)
Does continuous non-invasive cardiac output monitoring using radial artery applanation tonometry agree with invasive thermodilution and pulse contour analysis in ICU patients after major gastrointestinal surgery?
Mean Difference: -0.8 (95% CI -4–2.3)
Non-invasive cardiac output monitoring using the T-Line system showed high percentage errors compared to invasive methods, indicating further improvements are needed before clinical use in ICU patients.
TL-CO monitoring cannot replace invasive methods in post-surgical ICU patients; leaves open whether device refinements can achieve acceptable agreement.
The T-Line(®) system (Tensys(®) Medical Inc., San Diego, CA, USA) non-invasively estimates cardiac output (CO) using autocalibrating pulse contour analysis of the radial artery applanation tonometry-derived arterial waveform. We compared T-Line CO measurements (TL-CO) with invasively obtained CO measurements using transpulmonary thermodilution (TDCO) and calibrated pulse contour analysis (PC-CO) in patients after major gastrointestinal surgery. We compared 1) TL-CO versus TD-CO and 2) TL-CO versus PC-CO in 27 patients treated in the intensive care unit (ICU) after major gastrointestinal surgery. For the assessment of TD-CO and PC-CO we used the PiCCO(®) system (Pulsion Medical Systems SE, Feldkirchen, Germany). Per patient, we compared two sets of TD-CO and 30 minutes of PC-CO measurements with the simultaneously recorded TL-CO values using Bland-Altman analysis. The mean of differences (± standard deviation; 95% limits of agreement) between TL-CO and TD-CO was -0.8 (±1.6; -4.0 to +2.3) l/minute with a percentage error of 45%. For TL-CO versus PC-CO, we observed a mean of differences of -0.4 (±1.5; -3.4 to +2.5) l/minute with a percentage error of 43%. In ICU patients after major gastrointestinal surgery, continuous non-invasive CO measurement based on autocalibrating pulse contour analysis of the radial artery applanation tonometry-derived arterial waveform (TL-CO) is feasible in a clinical study setting. However, the agreement of TL-CO with TD-CO and PC-CO observed in our study indicates that further improvements are needed before the technology can be recommended for clinical use in these patients.
No takes yet. Share an insight, caveat, or question.
Wagner et al. (2016) conducted an observational in ICU patients after major gastrointestinal surgery (n=27). T-Line system (TL-CO) vs. Transpulmonary thermodilution (TD-CO) and calibrated pulse contour analysis (PC-CO) was evaluated on Agreement between TL-CO and TD-CO (MD -0.8 l/minute, 95% CI -4.0 to 2.3). Non-invasive cardiac output measurement using the T-Line system showed a mean difference of -0.8 l/minute (95% limits of agreement -4.0 to 2.3) compared to transpulmonary thermodilution.