Key result
Trifecta valve implantation demonstrated a 7-year freedom from structural valve deterioration rate of 92.6%, with 6.4% of patients experiencing early failure primarily due to noncoronary cusp tear.
Why the study?
Several studies have reported a high structural valve deterioration rate for the Trifecta valve, prompting an analysis of its midterm outcomes and risk factors for early deterioration.
Does the 19 mm Trifecta valve increase the risk of structural valve deterioration compared to other sizes in patients undergoing aortic valve replacement?
Observational (n=110)
No
Does the 19 mm Trifecta valve increase the risk of structural valve deterioration compared to other sizes in patients undergoing aortic valve replacement?
Small (19 mm) Trifecta valves and high postoperative mean pressure gradients may be risk factors for early structural valve deterioration, most commonly caused by noncoronary cusp tears.
Answers are grounded in this paper and cite their sources.
Trifecta recipients may need closer SVD surveillance; leaves open durability questions versus other bioprostheses in larger cohorts.
Background: Several studies have reported a high structural valve deterioration (SVD) rate for the Trifecta valve. We analyzed the midterm outcomes of the Trifecta valve and risk factors for early SVD.Methods: We retrospectively reviewed 110 patients who underwent Trifecta implantation between January 2012 and December 2017.Results: We encountered seven cases of Trifecta valve failure. We performed redo aortic valve replacements in five patients and transcatheter aortic valve replacements in two patients. The free rate of SVD was 96.1% at five years and 92.6% at seven years. The mean pressure gradient and peak velocity in the first postoperative echocardiogram in patients with SVD were higher than those in patients with SVD. The SVD-free rates of the 19 mm valve and other valves were 93% and 95.6% at five years and 89.6% vs. 95.6% at seven years, respectively. Although not statistically significant, the 19 mm valves tended to be SVD. Noncoronary cusp tears were seen in all patients who underwent a redo surgery.Conclusions: The most common cause of SVD was noncoronary cusp tear. Patients with a high postoperative mean pressure gradient are at a high risk of SVD. The small size of the Trifecta for a small annulus or Valsalva sinus may be a risk factor for early SVD.
No takes yet. Share an insight, caveat, or question.
Wakami et al. (2022) conducted an observational in Aortic valve disease (n=110). Trifecta valve was evaluated on Structural valve deterioration (SVD). Trifecta valve implantation demonstrated a 7-year freedom from structural valve deterioration rate of 92.6%, with 6.4% of patients experiencing early failure primarily due to noncoronary cusp tear.
Synapse has enriched 8 closely related papers on similar clinical questions. Consider them for comparative context: