Key result
Excessive pannus growth on a bovine pericardial valve explanted after more than 20 years may have protected the bioprosthesis from earlier failure.
Case Report (n=1)
Excessive pannus growth may paradoxically protect bioprosthetic heart valves from structural degeneration, contributing to exceptional longevity.
Answers are grounded in this paper and cite their sources.
Case report does not support practice changes; leaves open whether pannus may extend bioprosthetic durability.
As the population ages, bioprosthetic heart valves are increasingly being used to replace diseased native valves. Bioprosthetic valve durability depends on patient age and other factors, but rarely exceeds 15 years. Explanted bioprosthetic valves commonly show tissue degeneration, tears, and calcification. Host tissue overgrowth (pannus), to the extent of interfering with their function, is another finding in bioprostheses that have been in place for long periods. We present a case in which a bovine pericardial valve was explanted after more than 20 years of implantation. The longevity of this pericardial valve may have been related to excessive pannus growth, which most likely protected the valve from earlier failure.
No takes yet. Share an insight, caveat, or question.
Butany et al. (2006) conducted a case report in Bioprosthetic heart valve degeneration (n=1). Host tissue overgrowth (pannus) was evaluated on Valve longevity and failure. Excessive pannus growth on a bovine pericardial valve explanted after more than 20 years may have protected the bioprosthesis from earlier failure.