Key result
Small (19 mm) bileaflet and pericardial aortic valve prostheses showed no significant differences in resting hemodynamics, with 46% of bioprostheses showing non-significant central regurgitation.
Why the study?
Do small (19 mm) bileaflet aortic valve prostheses differ from pericardial bioprostheses in resting hemodynamics among patients with small aortic roots?
Observational (n=43)
Do small (19 mm) bileaflet aortic valve prostheses differ from pericardial bioprostheses in resting hemodynamics among patients with small aortic roots?
Both 19 mm bileaflet and pericardial prostheses demonstrate satisfactory resting hemodynamics and are suitable for implantation in small aortic roots.
Answers are grounded in this paper and cite their sources.
Comparable resting hemodynamics favor neither prosthesis in small roots; leaves open durability and exercise data.
The resting haemodynamics of five types of small (19 mm) aortic valve prosthesis (2 bileaflet, 3 pericardial) were evaluated with Doppler echocardiography in 43 patients. Two received St Jude Medical and six CarboMedics bileaflet valves and 35 were given bioprostheses--16 Ionescu-Shiley, four Mitroflow and 15 Labcor-Santiago. No significant differences in peak or mean transvalvular pressure drop or in effective valve area were found between the bileaflet and the pericardial valves or among the three types of bioprosthesis. All but one of the bileaflet prostheses showed a characteristic regurgitation pattern, with two lateral and one central jet, and 16 (46%) of the bioprostheses showed central regurgitation, but in no case were these jets haemodynamically significant. Thus the 19 mm bileaflet and the studied pericardial prostheses all have satisfactory resting haemodynamics, and all are suitable for implanting in small aortic roots.
No takes yet. Share an insight, caveat, or question.
González‐Juanatey et al. (1995) conducted an observational in Aortic valve replacement with small aortic root (n=43). 19 mm bileaflet aortic valve prostheses vs. 19 mm pericardial aortic valve prostheses was evaluated on Peak or mean transvalvular pressure drop and effective valve area. Small (19 mm) bileaflet and pericardial aortic valve prostheses showed no significant differences in resting hemodynamics, with 46% of bioprostheses showing non-significant central regurgitation.
Synapse has enriched 10 related papers. Consider them for comparative context: