Key result
The paper describes the engineering and development of an extracorporeal maglev centrifugal-type blood pump employing a permanent magnet-free disposable pump head.
A novel maglev centrifugal blood pump demonstrated stable levitation and adequate flow characteristics in a mock circulatory loop, suggesting potential for ventricular assist applications.
Answers are grounded in this paper and cite their sources.
Hypothesis-generating for magnet-free maglev pumps; requires in vivo validation before clinical consideration.
In the field of centrifugal blood pumps, contactless support of the impeller by a magnetic bearing has been identified as a promising method for reducing blood damage and to enhance durability. We developed a novel blood pump both with a two-degree-of-freedom radial controlled magnetic bearing and a contactless torque transmission mechanism placed in the outer side of a permanent magnet-free impeller. The proposed mechanism can realize not only a smooth secondary flow in the pump, but also a cost effective pump head. The prototype magnetic bearing and magnetic coupling succeeded in the stable levitation and contactless rotation of the impeller both in air and water. The pump can provide a flow rate of 5L/min against a head pressure of more than 350 mmHg. In order to simulate the ventricular assist condition, the prototype pump was connected to a mock systemic circulatory loop including a pulsatile pump. The dynamic stability of the impeller was experimentally verified when the pulsatile pressure was applied during the left ventricular assist.
No takes yet. Share an insight, caveat, or question.
MOMOSE et al. (2014) studied Extracorporeal circulation (engineering study). Extracorporeal maglev centrifugal-type blood pump was evaluated. The paper describes the engineering and development of an extracorporeal maglev centrifugal-type blood pump employing a permanent magnet-free disposable pump head.
Synapse has enriched 10 related papers. Consider them for comparative context: