Key result
In 3 successful 2-week implantations out of 13, the MedTech Dispo pump produced constant flow without thromboembolism or thrombus formation, maintaining low plasma-free hemoglobin.
Why the study?
Does the MedTech Dispo pump demonstrate sufficient biocompatibility and safety for 2-week circulatory support in a calf model?
Does the MedTech Dispo pump demonstrate sufficient biocompatibility and safety for 2-week circulatory support in a calf model?
The MedTech Dispo pump demonstrated sufficient performance and biocompatibility in a calf model to support its potential for short-term circulatory support.
Supports preclinical viability of the MedTech Dispo pump; leaves open human translation for short-term circulatory support.
Last year, we reported in vitro pump performance, low hemolytic characteristics, and initial in vivo evaluation of a disposable, magnetically levitated centrifugal blood pump, MedTech Dispo. As the first phase of the two-stage in vivo studies, in this study we have carried out a 2-week in vivo evaluation in calves. Male Holstein calves with body weight of 62.4–92.2 kg were used. Under general anesthesia, a left heart bypass with a MedTech Dispo pump was instituted between the left atrium and the descending aorta via left thoracotomy. Blood-contacting surface of the pump was coated with a 2-methacryloyloxyethyl phosphorylcholine polymer. Post-operatively, with activated clotting time controlled at 180–220 s using heparin and bypass flow rate maintained at 50 mL/kg/min, plasma-free hemoglobin (Hb), coagulation, and major organ functions were analyzed for evaluation of biocompatibility. The animals were electively sacrificed at the completion of the 2-week study to evaluate presence of thrombus inside the pump,together with an examination of major organs. To date, we have done 13 MedTech Dispo implantations, of which three went successfully for a 2-week duration. In these three cases, the pump produced a fairly constant flow of 50 mL/Kg/min. Neurological disorders and any symptoms of thromboembolism were not seen. Levels of plasma-free Hb were maintained very low. Major organ functions remained within normal ranges. Autopsy results revealed no thrombus formation inside the pump. In the last six cases, calves suffered from severe pneumonia and they were excluded from the analysis. The MedTech Dispo pump demonstrated sufficient pump performance and biocompatibility to meet requirements for 1-week circulatory support. The second phase (2-month in vivo study) is under way to prove the safety and efficacy of MedTech Dispo for 1-month applications.
No takes yet. Share an insight, caveat, or question.
Nagaoka et al. (2010) studied Bridge-to-bridge applications (n=13). MedTech Dispo pump was evaluated on Pump performance, biocompatibility, and presence of thrombus. In 3 successful 2-week implantations out of 13, the MedTech Dispo pump produced constant flow without thromboembolism or thrombus formation, maintaining low plasma-free hemoglobin.