Key result
The Abbott ADV cardiac troponin I assay exhibited significantly higher relative concentrations from 30 minutes after admission compared to cTnT, CKMBmass, myoglobin, and H-FABP (P<0.05).
Why the study?
Does the Abbott ADV cTnI assay exhibit equal or better sensitivity compared to cTnT and other markers of myocardial necrosis in patients with acute MI?
Observational (n=23)
Does the Abbott ADV cTnI assay exhibit equal or better sensitivity compared to cTnT and other markers of myocardial necrosis in patients with acute MI?
p-value: p=<0.05
The Abbott ADV cTnI assay demonstrates higher relative concentrations and potentially greater early sensitivity than cTnT and other necrosis biomarkers in the first 24 hours of STEMI.
OBJECTIVE: Highly sensitive and specific assays of cardiac troponins I and T are the preferred biomarkers in diagnosing myocardial infarction (MI). Assays of cardiac troponin I (cTnI) have been improved with the addition of antibodies against the cTnI molecule and may have increased sensitivity. We hypothesized that a cTnI assay with modern antibody configuration will exhibit equal or better sensitivity in the setting of acute MI compared to cTnT and other markers of myocardial necrosis. MATERIAL AND METHODS: We investigated release kinetics of cTnI (Abbott ADV, Abbott Diagnostics), cTnT (Roche Diagnostics), CKMBmass, myoglobin and heart fatty acid binding protein (H-FABP) in 23 patients admitted with acute ST-segment elevation MI undergoing primary percutaneous coronary intervention. Calibrators for the Abbott ADV cTnI assay are traceable to the United States National Institute of Standards and Technology (NIST) reference material for cTnI. Eleven blood samples were drawn from each patient in the period from admission to 24 h. Biomarkers of necrosis showed marked increases in relative concentrations, especially within the first 2 h after admission. RESULTS: From 30 min after admission onwards, cTnI exhibited significantly higher relative concentrations compared to cTnT, CKMBmass, Myoglobin and H-FABP (p<0.05). CONCLUSIONS: The NIST standardized Abbott TnI ADV assay appears to be more sensitive than cTnT and other biomarkers in the early phase of MI.
No takes yet. Share an insight, caveat, or question.
Hjortshøj et al. (2008) conducted an observational in Acute ST-segment elevation myocardial infarction (n=23). Abbott ADV cardiac troponin I assay vs. cTnT, CKMBmass, myoglobin, and H-FABP assays was evaluated on Relative concentrations of biomarkers of necrosis (p=<0.05). The Abbott ADV cardiac troponin I assay exhibited significantly higher relative concentrations from 30 minutes after admission compared to cTnT, CKMBmass, myoglobin, and H-FABP (P<0.05).
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: