Key result
Crimping of textile valve prototypes generated creases and slightly reduced the mechanical properties of the materials, with variations depending on fabric construction and crimping configuration.
Crimping slightly reduces the mechanical properties of textile polyester materials proposed for TAVI leaflets, highlighting the need for further long-term durability testing.
Transcatheter aortic valve implantation (TAVI) has become a popular alternative technique to surgical valve replacement. However, the biological valve tissue used in these devices appears to be fragile material in the long term particularly due being folded for low diameter catheter insertion purposes and when released in a calcified environment with irregular geometry. Textile polyester material is characterized by outstanding folding and strength properties combined with proven biocompatibility. It could therefore be considered as a replacement for biological valve leaflets in the TAVI procedure. The folding process associated with crimping, however, may degrade the filaments involved in the fibrous assembly and limit the durability of the device. The purpose of the present work is to study the effect of different crimping conditions on the mechanical performances of textile valve prototypes made from various fabric constructions. Results show that crimping generates some creases in the fabrics, which surface topography varies with fabric construction and crimping configuration. The mechanical properties of the crimped materials are globally slightly reduced. To determine how critical the modifications due to crimping are for prosthesis durability, more detailed long term in vitro and in vivo trials with crimped textile prototypes are needed in addition to this preliminary work.
No takes yet. Share an insight, caveat, or question.
Khoffi et al. (2014) studied Transcatheter aortic valve implantation. Crimping vs. Different crimping conditions was evaluated on Mechanical performances. Crimping of textile valve prototypes generated creases and slightly reduced the mechanical properties of the materials, with variations depending on fabric construction and crimping configuration.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: