Key result
Portable coagulometers provided INR results comparable to laboratory tests, with CoaguChek XS Plus showing higher concordance than INRatio PT Monitor.
Why the study?
Do portable coagulometers provide comparable INR results to laboratory venipuncture in patients on Vitamin K antagonists?
Observational (n=1,009)
No
Do portable coagulometers provide comparable INR results to laboratory venipuncture in patients on Vitamin K antagonists?
Answers are grounded in this paper and cite their sources.
Portable coagulometers, particularly the CoaguChek XS Plus, provide reliable INR measurements that correlate well with laboratory venipuncture, especially for values within or below the therapeutic range.
INTRODUCTION: Therapy with Vitamin K antagonists (VKA) is effective in reducing thromboembolic events in various diseases. There are limitations however, which limit clinical handling and maintaining INR within therapeutic range. Studies have shown that portable coagulometers, when compared to laboratory tests, are more practical and provide better patient adherence and involvement toward treatment which results in better INR control. This study aimed to evaluate laboratory obtained INR results compared to two different portable coagulometers. METHODS: A prospective study which monitored 1009 patients using VKA in the Anticoagulation Clinic at the Institute Dante Pazzanese of Cardiology in São Paulo between July and September 2012. Patient INR values were obtained by the laboratory through venipuncture and then compared to INR values obtained by capillary puncture from two different portable coagulometers. RESULTS: Overall, 1009 patients were included in the study; among these, 520 (51.5%) are male with average age of 59.6 years (13-91). The more common indications were atrial fibrillation (49.9%) and mechanical prosthesis (33.7%). The correlation coefficient was of 0.95 with and 0.88 with INRatio PT Monitor(®) compared to laboratory. In patients with INR < 2 (lower than therapeutic range), the coefficient was 0.92 and 0.81 for CoaguChek XS plus(®) and INRatio PT Monitor(®) respectively. In patients within therapeutic range (INR 2-3), the coefficient was 0.86 with CoaguChek XS Plus(®) and 0.76 with INRatio PT Monitor(®) . For INR above therapeutic range (INR > 3.0) the correlation was 0.80 with CoaguChek XS Plus(®) and 0.54 with INRatio PT Monitor(®) . As for concordance between methods, the intraclass correlation coefficients (ICC) were slightly smaller than those previously stated (ICC = 0.899 with CoaguChek XS Plus(®) and ICC = 0.716 with INRatio PT Monitor(®) ). CONCLUSION: The use of portable coagulometers was comparable to laboratory tests and better correlation coefficients were observed with CoaguChek XS Plus(®) and in patients with INR lower or within therapeutic range. Portable coagulometers proved to be a useful and reliable tool for INR control in patients using VKA.
No takes yet. Share an insight, caveat, or question.
Meneghelo et al. (2015) conducted an observational in Patients using Vitamin K antagonists (n=1,009). Portable coagulometers (CoaguChek XS Plus and INRatio PT Monitor) vs. Laboratory tests (venipuncture) was evaluated on Correlation and concordance of INR results between portable coagulometers and laboratory tests (ICC 0.899 (CoaguChek XS Plus) and 0.716 (INRatio PT Monitor)). Portable coagulometers provided INR results comparable to laboratory tests, with CoaguChek XS Plus showing higher concordance than INRatio PT Monitor.